#!/usr/bin/env python3
"""
RION NL-42 / NL-52 USB serial probe — zero dependencies (stdlib termios only).
The NL-52's USB port enumerates as a virtual COM port ("RION USB to RS232C
Converter"). On Linux that is almost always handled by an in-kernel USB-serial
driver (ftdi_sio / cp210x / ch341) which creates /dev/ttyUSB*. This script
opens that port and sends a few harmless REQUEST commands (no settings are
changed) to confirm two-way communication.
Protocol (NL-42/NL-52 Serial Interface Manual 55779):
Request: "<Command>?" + CRLF
Setting: "$<Command>,<param>" + CRLF (NOT used here — read-only probe)
Reply: result code "R+0000" + CRLF, then data line(s) for requests.
Before running:
1. On the meter: MENU -> I/O -> Communication Interface -> "USB"
(set this BEFORE plugging in the cable).
2. Connect a generic USB-A -> mini-B cable directly (no hub).
3. Find the port: ls -l /dev/ttyUSB* (and `dmesg | tail` after plugging in)
Usage:
python3 nl52_usb_probe.py # defaults to /dev/ttyUSB0
python3 nl52_usb_probe.py --port /dev/ttyUSB0 --baud 115200
python3 nl52_usb_probe.py --cmd "System Version?" --cmd "Clock?"
"""
import argparse
import os
import select
import sys
import termios
import time
# Safe, read-only probe commands (all are pure requests).
DEFAULT_COMMANDS = [
"System Version?", # firmware version — proves the link end-to-end
"Clock?", # current date/time
"SD Card Free Size?",
"DOD?", # snapshot of currently displayed values
]
BAUD_CONSTANTS = {
9600: termios.B9600,
19200: termios.B19200,
38400: termios.B38400,
57600: termios.B57600,
115200: termios.B115200,
}
def open_serial(port: str, baud: int) -> int:
"""Open a serial port in raw 8N1, no flow control. Returns an fd."""
if baud not in BAUD_CONSTANTS:
raise ValueError(f"Unsupported baud {baud}; choose from {sorted(BAUD_CONSTANTS)}")
fd = os.open(port, os.O_RDWR | os.O_NOCTTY | os.O_NONBLOCK)
attrs = termios.tcgetattr(fd)
iflag, oflag, cflag, lflag, ispeed, ospeed, cc = attrs
# Raw mode
iflag = 0
oflag = 0
lflag = 0
# 8 data bits, enable receiver, ignore modem control lines
cflag = termios.CS8 | termios.CREAD | termios.CLOCAL
# (no PARENB = no parity, no CSTOPB = 1 stop bit, no CRTSCTS = no flow control)
bconst = BAUD_CONSTANTS[baud]
ispeed = bconst
ospeed = bconst
termios.tcsetattr(fd, termios.TCSANOW, [iflag, oflag, cflag, lflag, ispeed, ospeed, cc])
termios.tcflush(fd, termios.TCIOFLUSH)
return fd
def send(fd: int, line: str):
os.write(fd, (line + "\r\n").encode("ascii"))
def read_reply(fd: int, timeout: float = 3.0) -> bytes:
"""Read whatever arrives within `timeout` seconds (idle-gap terminated)."""
buf = bytearray()
deadline = time.time() + timeout
while time.time() < deadline:
r, _, _ = select.select([fd], [], [], 0.3)
if r:
try:
chunk = os.read(fd, 4096)
except BlockingIOError:
continue
if chunk:
buf.extend(chunk)
# Once we've seen a CRLF and there's a brief idle, stop early
deadline = min(deadline, time.time() + 0.4)
return bytes(buf)
def main():
ap = argparse.ArgumentParser(description="RION NL-42/NL-52 USB serial probe")
ap.add_argument("--port", default="/dev/ttyUSB0")
ap.add_argument("--baud", type=int, default=115200,
help="USB CDC usually ignores baud, but RS-232C needs a match")
ap.add_argument("--cmd", action="append", dest="cmds",
help="Override probe command(s); repeatable")
ap.add_argument("--timeout", type=float, default=3.0)
args = ap.parse_args()
commands = args.cmds or DEFAULT_COMMANDS
if not os.path.exists(args.port):
print(f"[!] {args.port} does not exist.")
print(" Plug in the meter (Comm Interface = USB) and check: ls -l /dev/ttyUSB*")
print(" Also check the kernel saw it: dmesg | tail -20")
return 2
try:
fd = open_serial(args.port, args.baud)
except PermissionError:
print(f"[!] Permission denied on {args.port}.")
print(" Add yourself to the 'dialout' group, or run with sudo:")
print(f" sudo usermod -aG dialout $USER (then log out/in)")
return 2
except Exception as e:
print(f"[!] Could not open {args.port}: {e}")
return 2
print(f"[*] Opened {args.port} @ {args.baud} 8N1 (raw, no flow control)")
print(f"[*] Sending {len(commands)} read-only request command(s)\n")
ok = 0
try:
for cmd in commands:
send(fd, cmd)
reply = read_reply(fd, args.timeout)
decoded = reply.decode("ascii", errors="replace").replace("\r", "\\r").replace("\n", "\\n\n")
if reply:
ok += 1
print(f" > {cmd}")
for line in decoded.splitlines():
print(f" {line}")
else:
print(f" > {cmd}")
print(f" (no response within {args.timeout}s)")
print()
time.sleep(1.0) # NL-series likes >=1s between commands
finally:
os.close(fd)
if ok == 0:
print("[!] No responses. Things to check:")
print(" - Meter's Communication Interface is set to USB (not RS-232C)")
print(" - ECO / Sleep mode is OFF (both disable the comm interface)")
print(" - Right port (try other /dev/ttyUSB* or /dev/ttyACM*)")
print(" - For RS-232C path, baud must match the meter's setting")
return 1
print(f"[OK] {ok}/{len(commands)} commands answered — two-way comms confirmed.")
return 0
if __name__ == "__main__":
sys.exit(main())