Brian's intent (UM6047's trigger, notes typos, the backup unit), the flow, the first-version fields, and what is measured: Series 3's compliance block has Series 4's 0x1A layout; a whole 1A read assembles to the 2,090 bytes Blastware writes back (exact on both 10-26 writes); Blastware writes 00 00 at [26:28] and a per-unit constant after the block. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
This commit is contained in:
1 parent
cbce00f8e1
commit
0238e4d3f3
1 file changed
+339
@@ -0,0 +1,339 @@
|
||||
# Phase 5: changing a unit's settings by serial (design)
|
||||
|
||||
**Date:** 2026-11-03. **Status:** draft for Brian. Phase 5 of
|
||||
`docs/superpowers/specs/2026-10-20-sfm-series4-parity-design.md` (row 10, "change
|
||||
recording settings"; build phase 5, "setup push: one field first, then several").
|
||||
**Branch:** `feat/series4-sfm-parity`.
|
||||
|
||||
## Intent (Brian, 2026-11-03)
|
||||
|
||||
Change **any** of a unit's settings from SFM, and make it as easy and reliable as
|
||||
possible. The goal is to fix what makes Blastware finicky. Three real cases:
|
||||
|
||||
- **One wrong number on a deployed unit.** UM6047 was out monitoring with its
|
||||
geophone trigger at 0.03 instead of 0.300. Brian remoted into the THOR PC,
|
||||
stopped it (it called in), changed the setup, made sure it was *sent*, then
|
||||
restarted monitoring.
|
||||
- **A typo in the notes.**
|
||||
- **A backup unit.** The unit set up at the office fails in the field, so a backup
|
||||
from the truck goes in with blank or stale notes and settings. There is no
|
||||
computer in the field, so the office re-does the whole setup remotely, by hand,
|
||||
while the person in the field stands by to press Start or power-cycle if the
|
||||
link breaks.
|
||||
|
||||
What is wrong with THOR, as Brian and his dad experience it:
|
||||
|
||||
- The same screen edits a setup on disk and views a unit's setup.
|
||||
- Setups get saved under random number strings.
|
||||
- Editing a setup does not change the unit; you verify separately that it was sent.
|
||||
|
||||
Saved setups as templates would be acceptable. Not knowing whether you are
|
||||
changing the unit or a file is not.
|
||||
|
||||
## Goals, non-goals, anti-goals
|
||||
|
||||
**Goals**
|
||||
|
||||
- Edit the unit's **live** settings, both series.
|
||||
- Show exactly what will change, **old → new**, before anything is sent.
|
||||
- After sending, **re-read the unit** and show what it now holds.
|
||||
- Make every refusal say why, in plain words.
|
||||
|
||||
**Non-goals (this phase)**
|
||||
|
||||
- Templates, saved setups, copying from another unit.
|
||||
- Restarting monitoring. **Brian presses Start**, as today; SFM never restarts on
|
||||
its own.
|
||||
- Setting the unit's clock. That is a later call-home receiver rule.
|
||||
- Fields not yet decoded. "Any and all" grows as decoding grows.
|
||||
|
||||
**Anti-goals: things that would make it a failure even if it "works"**
|
||||
|
||||
- Any doubt about whether the unit changed.
|
||||
- A screen where a file and a unit look alike, or a name SFM made up.
|
||||
- Writing a byte SFM did not just read. Every byte SFM does not deliberately change
|
||||
goes back exactly as read. The one exception is a byte that Blastware's (or
|
||||
THOR's) own writes always set to a fixed value. Each such byte is named in the
|
||||
codec with its evidence (so far: Series 3 `[26:28]` → `00 00`; my call, below).
|
||||
- Sending to a monitoring unit.
|
||||
- A retry. A write is sent once; what happened is read back.
|
||||
|
||||
## Measured basis
|
||||
|
||||
**1. Series 4 (Micromate): the `0x1A` setup content map.** Measured from 47
|
||||
keypad snapshots on UM20147 and checksum-verified. `micromate/setup_block.py`
|
||||
(`decode_setup`) reads it, with 69 tests on 15 fixtures.
|
||||
|
||||
| content | field |
|
||||
|---|---|
|
||||
| `[5]` | record mode |
|
||||
| `[6:8]` | sample rate |
|
||||
| `[8:10]` | histogram interval |
|
||||
| `[12:16]` | pre-trigger, f32, negative |
|
||||
| `[17]` | stop mode |
|
||||
| `[18:22]` | record time, f32 |
|
||||
| `[23]` | auto-stop |
|
||||
| `[25]` | active sensors |
|
||||
| `[29]` | trigger sources |
|
||||
| `[30:34]` | distance |
|
||||
| `[34:38]` | charge |
|
||||
| `[39]` | report flags |
|
||||
| `[40:42]` | job number |
|
||||
| `[42:86]` | setup name |
|
||||
| `[122+64n]` / `[144+64n]` | notes (label / value) |
|
||||
| from `[1708]` | 48-byte channel blocks: sensor code `+4:8`, trigger `+30` (geo in/s, mic psi) |
|
||||
|
||||
The CRC-16/AUG-CCITT over `[0:2088]` is stored at `[2090:2092]`; THOR's write
|
||||
omits it and the unit recomputes it.
|
||||
|
||||
- **The push sequence (from THOR's captures)** is `0xDA` (name) → `0x68`/`0x73`
|
||||
(`0x08` preferences) → `0x82`/`0x83` (`0x2E` trigger) → `0x71`/`0x72` (`0x1A`
|
||||
setup). `micromate.writes.plan_setup_push` reproduces THOR's frames (phase 2).
|
||||
- **Not yet mapped on Series 4:** alarm level, geophone range, GPS, aux I/O, SLM
|
||||
options.
|
||||
|
||||
**2. Series 3 (MiniMate) uses the same layout. Measured 2026-11-03.** The
|
||||
compliance image as Blastware writes it is logical (unescaped), 2,092 bytes, with
|
||||
`N = 2090` at `[0:2]` and the channel block "Tran" at `[1708]`.
|
||||
`micromate.setup_block.decode_setup` reads it correctly:
|
||||
|
||||
- **BE9441's 10-26 write** (`tests/fixtures/minimate-bench-10-26-26-histogram`,
|
||||
two whole sends): name, mode, 4096 sps, interval, record time 20 s, trigger
|
||||
0.04 in/s, mic linear, job number and the notes all decode. Only the CRC fails
|
||||
(Series 3's tail goes out stale).
|
||||
- **BE11529's nine 2026-04-20 Blastware writes** (`bridges/captures/4-20-26/`,
|
||||
every one with a whole image) each changed exactly the expected bytes:
|
||||
|
||||
| Blastware changed | bytes that moved |
|
||||
|---|---|
|
||||
| histogram interval (×4) | `[8:10]` |
|
||||
| recording mode (×3) | `[5]` |
|
||||
| geophone range | `1737` / `1785` / `1833`, the range byte at each geophone channel label `+29` |
|
||||
|
||||
- **The live `SUB 1A` read starts the same block.** Measured on the first data
|
||||
frame of each of the four recorded reads (`_settings_replies` in
|
||||
`tests/test_s3_compliance_anchor_fields.py`): unescaped, each carries `N = 2090`
|
||||
12 bytes before the old "anchor", and content `[5]` reads 1 / 4 / 3 / 3. That
|
||||
anchor (`be 80 00 00 00 00`) is content `[12:18]`, pre-trigger −0.25 plus stop
|
||||
mode Fixed. So Series 3 can be read by content offset rather than by anchor
|
||||
search, which survives an Auto-stop unit (the open question of 2026-10-17).
|
||||
- **A whole `1A` read assembles to exactly the 2,090 content bytes** (measured
|
||||
2026-11-03 on BE9441's 10-26 recording, 12 reads). Unescape each data frame and
|
||||
take `data[11:]`. Every frame but the last then starts with one `00` byte that is
|
||||
not content; drop it. This holds for Blastware's read (four 512-byte frames plus
|
||||
42) and SFM's (two 1024-byte frames plus 42).
|
||||
- **The round trip is exact.** Blastware re-read the block after each of its two
|
||||
writes, and each re-read equals the image it had just written, all 2,090 bytes.
|
||||
- **While monitoring**, the read returns `0e 'BE9441' …` (the device-info block),
|
||||
as the 10-29 finding says. The unit was monitoring for 5 of the 12 reads.
|
||||
- **Bytes `[26:28]`: Blastware always writes `00 00`** (all 11 writes), though the
|
||||
unit reads `01 7f` after a monitoring session (BE9441 10-26, BE11529 4-20 15:19).
|
||||
The unit sets them again by itself. Meaning unknown.
|
||||
- **The two bytes after the block are Blastware's own.** Blastware sends a per-unit
|
||||
constant: `7c ea` on all nine of BE11529's writes, `0c e9` on both of BE9441's.
|
||||
It is not the block's CRC-16/AUG, and no reply in BE9441's three-hour session
|
||||
contains it. SFM's `write_image` sends `00 00` there. Since a read never returns
|
||||
those bytes, the read-back cannot show what the unit did with them.
|
||||
- **The mode byte is shared, but its codes are not.** Series 3: `00` Single Shot,
|
||||
`01` Continuous, `03` Histogram, `04` Histogram + Continuous. Series 4: `01`
|
||||
Waveform, `02` Manual, `03` Histogram, `04` Combo.
|
||||
- **Rules from the checker's file-side work** (seismo-relay-b6, 2026-11-03; it
|
||||
measured the same content offsets on 70,042 event files and found nothing that
|
||||
contradicts one shared map):
|
||||
- A bitmask byte is patched **bit by bit, never rebuilt**. The trigger-sources
|
||||
byte `[29]` showed an unidentified `0x80` bit on one unit.
|
||||
- On Series 3, stop mode (`0` Fixed) and mic type (`2` linear) are proven for
|
||||
one value each. They are not offered for writing.
|
||||
- An event file's `[12:16]` is that event's captured pre-trigger, not the
|
||||
setting. SFM writes only from a live read, so this matters only to anyone
|
||||
tempted to build a write from a file.
|
||||
- **Series-3-only fields** (measured on Series 3, not Series 4): the geophone range
|
||||
byte at channel `+29`; the alarm level at channel `+38` (checker-measured on
|
||||
event files; live raw `+42`).
|
||||
|
||||
**3. Series 3 write frames are measured (#46).** Writes use uniform escaping with
|
||||
a length field. The compliance image goes in 512-byte addressed chunks. `0x68` is
|
||||
written as read plus `02 00`, `0x82` plus `00 00`. The sequence is
|
||||
`68→73 | 71×n→72 | 82→83 | 69→74→72`. `push_config_raw` refuses any block that
|
||||
is not whole. `tests/test_minimate_write_frames.py` rebuilds Blastware's frames
|
||||
byte for byte.
|
||||
|
||||
**4. Monitoring.**
|
||||
|
||||
- **Series 3:** while monitoring, the unit answers `1A` (and `0x2C`) with its
|
||||
device-info block (measured 10-29 and 11-03).
|
||||
- **Series 4:** it refuses call-home setup while monitoring ("Feature disabled
|
||||
while monitoring", measured 11-03).
|
||||
- **A stop with call-home on makes the unit call in** (measured 10-25, 10-29,
|
||||
11-03).
|
||||
|
||||
**5. Series 4 specifics.**
|
||||
|
||||
- `0x1A` reads the keypad's edit buffer live: unsaved keypad edits show in the
|
||||
read, and a push would commit them.
|
||||
- Loading a `.MMB` replaces the active config byte for byte. The push writes
|
||||
under the active setup's name (`0xDA`).
|
||||
|
||||
## The flow
|
||||
|
||||
1. **Read.** `GET /device/settings?serial=…` (it exists) returns the unit's
|
||||
settings. The page shows them on the Config tab under **"On the unit BE9441 —
|
||||
read 14:03"**, never anything that looks like a file.
|
||||
2. **Edit and preview.** The operator changes fields, then **Preview**: a dry run.
|
||||
SFM re-reads the unit's blocks on the same line, builds the new blocks, and
|
||||
returns one row per field: **old → new**, plus any warnings. Nothing is sent.
|
||||
3. **Send.** The same request, live, is an audited change:
|
||||
- identity check, audit row, one send, never retried, read back;
|
||||
- **refused with "stop monitoring first"** if the unit is monitoring, noting
|
||||
that the stop makes it call in when call-home is on;
|
||||
- **refused** if the unit's settings changed since the preview. The preview
|
||||
carries a revision (a hash of the blocks read), and the live call must
|
||||
match it.
|
||||
4. **Verify.** Re-read the blocks. Every changed field must read its new value.
|
||||
Every other decoded field, and every byte SFM did not deliberately change, must
|
||||
read as before. The page shows **"the unit now holds"**: field, old, new, now.
|
||||
Known unit-maintained bytes (the Series-4 CRC) are compared as the unit's own
|
||||
checksum (`crc_ok`), not byte for byte.
|
||||
5. **Start** stays the operator's button.
|
||||
|
||||
## Fields in the first version (Brian approved the list)
|
||||
|
||||
| field | Series 3 | Series 4 |
|
||||
|---|---|---|
|
||||
| geophone trigger level | yes (channel `+30`) | yes (channel `+30`) |
|
||||
| alarm level | yes (channel `+38`; measure live) | not mapped, so not offered |
|
||||
| notes: the four note slots, by the unit's own labels | yes (`Project:`, `Client:`, `User Name:`, `Seis. Loc:` on BE9441) | yes (`Location`, `Client`, `Company`, `General Notes`) |
|
||||
| recording mode | yes (`[5]`; values as measured on Series 3) | yes (`[5]`; Series-4 codes) |
|
||||
| sample rate | yes | yes |
|
||||
| record time | yes | yes |
|
||||
| histogram interval | yes | yes |
|
||||
| geophone range | yes (channel `+29`) | not mapped, so not offered |
|
||||
| job number (+ shown on report) | yes | yes |
|
||||
|
||||
A field is offered only where it is measured for that series. Values are refused
|
||||
unless they are among the measured choices: sample rates 1024/2048/4096,
|
||||
intervals 2/5/15/60/300/900 s, and each series' measured mode codes.
|
||||
|
||||
- **Numbers (my call; no unit limits are written down anywhere in the repo):**
|
||||
- Refuse any value outside the span seen on real units. The builder derives
|
||||
each span from the DL2 XML reports, the fixtures and the captures, and records
|
||||
it in the codec with its source. SFM then never sends a value no unit has
|
||||
been seen to hold. A span can be widened later on evidence.
|
||||
- A geophone trigger or alarm must also be at or below the range's full scale
|
||||
(10 / 1.25 in/s). When range and trigger change in one request, check the
|
||||
trigger against the **new** range.
|
||||
- **Units:** refuse a level write unless the channel's unit string reads the
|
||||
measured one (`in.` for a geophone). A unit set to Metric is unmeasured.
|
||||
- **Notes:** edited by slot, under the label the unit itself shows (labels can be
|
||||
renamed and are not identifiers). The value is at most its field width less
|
||||
one byte, printable ASCII.
|
||||
- Not in v1: Series 3's fifth slot, `Extended Notes` (value `\n` on BE9441).
|
||||
Blastware's extended notes are unmeasured and may live elsewhere.
|
||||
- Not in v1: Series 4's 16-byte field at content `[1064]` (`Loc 1` on
|
||||
UM12947), undecoded.
|
||||
|
||||
## Architecture
|
||||
|
||||
- **One setup-block codec for both series.** Keep `micromate/setup_block.py` as
|
||||
the field map and add `encode_setup(content, changes) -> content`: field patches
|
||||
only, every other byte as given. It stays a pure codec library: no I/O, no
|
||||
`sfm` imports, no import cycle with `minimateplus`. Series-specific fields (the
|
||||
Series-3 range and alarm) are marked by series.
|
||||
- **Series 3 read.** Assemble the `1A` read into the logical 2,090-byte block:
|
||||
unescape, strip the 12-byte reply prefix and the inter-frame headers
|
||||
(`read_compliance_config`). Decode it by content offset.
|
||||
- **Series 3 write.** The patched logical block plus the tail becomes the
|
||||
compliance image for `push_config_raw`. Event index (`0x08`), trigger (`0x2E`)
|
||||
and waveform data (`0x09`) are echoed as read, in Blastware's sequence.
|
||||
- On `0x09`, Blastware echoed the unit's bytes on all nine April writes
|
||||
(`[200:202]` = `8c 76` read and written). Only BE9441's 10-26 writes changed
|
||||
them (`17 17` read, `78 76` written, and read back so). The meaning is
|
||||
unknown, so SFM echoes.
|
||||
- **The raw-offset `_encode_compliance_config` is retired.**
|
||||
- **The legacy write (`/device/config`, `/device/config/project`) is retired (my
|
||||
call).** It retries a whole write on a TCP error, is not audited, does no
|
||||
read-back, does not refuse a monitoring unit, and needs no serial: every
|
||||
anti-goal at once. Its only user is the webapp's by-host Config form, and
|
||||
Terra-View does not call it. Its routes answer with a pointer to the by-serial
|
||||
change, and the by-host form becomes read-only.
|
||||
- **Series 4 write.** `plan_setup_push(prefs, trigger, patched_setup, name)`.
|
||||
- **No setup files in the operator's view (Brian, 2026-11-03).** "We treat the
|
||||
Micromate setups just like Minimate setups … the unit just has its settings,
|
||||
and we change it based on the job. We don't send setups to the unit." A
|
||||
backup unit is usually on whatever setup its previous job left. So SFM
|
||||
rewrites the setup the unit is running, in place, under its own name. It
|
||||
never asks for a file or a name, and never creates one. The name may show as
|
||||
a read-only detail, since the unit's screen shows it.
|
||||
- The name comes from the `0x41` read, unstripped.
|
||||
- Refuse `factory.MMB` and any target that is not the active setup (spec). The
|
||||
refusal says it plainly: "this unit is running its factory setup; load any
|
||||
other setup on its keypad, then try again." Brian says the team does not use
|
||||
the factory setup, so this should be rare.
|
||||
- `ENABLEABLE_WRITES` gains `0xDA 0x68 0x73 0x82 0x83 0x71 0x72`.
|
||||
- The keypad edit buffer: warn in the preview that a unit with someone in its
|
||||
menu would have those edits committed. The check the spec names, `0x1A`
|
||||
against the active `.MMB` via `0x94`/`0x48`, is a bench question first.
|
||||
- **The change path.** It keeps run_change's guarantees: identity, audit, one
|
||||
send, read back, `ChangeUnconfirmed` with `sent`/`audit_id`. The "state" is a
|
||||
set of fields, so either `run_change` takes a verify function or a sibling core
|
||||
shares its parts. The builder decides; the guarantees do not change. The audit
|
||||
row keeps the per-field changes, both blocks (hex) and the verdict.
|
||||
- **Routes.** `POST /device/settings/change?serial=…` (serial required) with body
|
||||
`{changes: {field: value}, revision}` and `dry_run`, `caller`. The response is
|
||||
per-field `old` / `new` / `now`, through `_change_response`. A dry run needs no
|
||||
revision and returns one. **A live call without the revision of a preview is
|
||||
refused** (my call): nothing reaches a unit unless someone saw old → new.
|
||||
- **Page.** The Config tab by serial becomes the unit's edit form: Preview → Send
|
||||
(confirm) → "the unit now holds". New logic goes in
|
||||
`sfm/static/device_routing.js` with Node tests, and glue in
|
||||
`sfm/static/sfm_webapp.js` (no inline JS).
|
||||
|
||||
## Verification
|
||||
|
||||
- **Series 3, byte for byte.** For each of the 11 Blastware compliance writes,
|
||||
apply the new encoder to the read before it (or the previous image), with the
|
||||
fields Blastware changed. It must reproduce Blastware's content `[0:2090]`. The
|
||||
two trailing bytes are compared separately, since they are Blastware's constant.
|
||||
Ten writes change one field. BE9441's 16:26 write changes five things: mode,
|
||||
sample rate 2048 → 4096, interval, the mic in the sensors byte `[25]`, and
|
||||
`[26:28]`. That one is tested with all of them, through the codec, though v1's
|
||||
page does not offer the sensors byte.
|
||||
The frames rebuild through `push_config_raw`'s path (the #46 tests).
|
||||
- **Series 3, read side.** The assembly rule above, pinned as a test on the 12
|
||||
reads of BE9441's 10-26 recording. Each read-back equals the write before it.
|
||||
- **Series 4.** THOR's setup-push captures still reproduce, with no field changed.
|
||||
Any capture where THOR changed a field becomes a byte-for-byte test.
|
||||
- **Bench, with Brian's go per send, per series:**
|
||||
1. A dry run of a trigger change, then send, read back, check the unit's menu,
|
||||
and change it back.
|
||||
2. A notes change.
|
||||
3. A several-field "backup unit" change.
|
||||
4. A dry run while monitoring, expecting the refusal.
|
||||
|
||||
## Calls I made (Claude), for Brian to check
|
||||
|
||||
- **Series 3 `[26:28]` goes out as `00 00`, as Blastware writes it**, rather than
|
||||
the `01 7f` the unit may read. Measured beats inferred: Blastware's value is on
|
||||
11 accepted writes, and writing back `01 7f` has never been tried.
|
||||
- **The two bytes after the Series 3 block stay `00 00`**, as `write_image` sends
|
||||
them (a unit accepted that in the April 2026 test). Blastware's per-unit
|
||||
constant cannot be derived from anything SFM reads. The first bench write checks
|
||||
that the unit takes the block, and its menu shows the change.
|
||||
- **The `0x09` block is echoed as read**, as Blastware did on all nine April writes.
|
||||
- **Numbers are bounded by what real units have been seen to hold**, and level
|
||||
writes are refused unless the channel's unit string is the measured `in.`.
|
||||
- **Notes are edited by slot under the unit's own labels**, four slots on each
|
||||
series. Series 3 extended notes and Series 4's separate location field wait.
|
||||
- **The legacy by-host write is retired**; the by-host form becomes read-only.
|
||||
- **A live send needs a preview's revision.**
|
||||
- **Preview re-reads the unit** rather than trusting the page's copy, and Send
|
||||
refuses if the unit changed since the preview.
|
||||
- **The edit form lives on the Config tab, by serial**, replacing the legacy
|
||||
read-modify-write form for units reached by serial.
|
||||
|
||||
## Left to the builder
|
||||
|
||||
- The module layout for the shared codec, within the no-cycle, pure-codec rule.
|
||||
- `run_change` generalised versus a sibling core.
|
||||
- The exact revision hash.
|
||||
- Page layout details.
|
||||