@@ -740,6 +740,20 @@ items were left for later on purpose:
## Parking lot
- **Speed (Brian's goal: faster than Blastware), after 1c.** Measured on
BE9441, 2026-10-27 23:41, serial cable:
- wake + serial number take 0.16 s;
- `MiniMateClient.connect()` also reads 01 + 1A + 08 on every request, ~3 s;
- a fixed **2.0 s "safety drain"** at the end of `read_compliance_config` runs
on every connect;
- the read-back polls every 2 s.
Ideas, each measured before and after against Blastware's own timings in the
bridge recordings: a light connect (wake + serial) for start/stop, status and
event list; drop or shorten the drain; reuse within a session (MiniMates
reconnect per request). New safeguards must state their time cost
(1c's identity check reuses the serial read connect already does).
**Monitor logs.** Its own piece of work, data side as much as device side.
A unit calls in at least once or twice a day. With no events (e.g. waveform
mode, no blast that day) it still delivers a **monitor log**: "monitoring from