Measured at the bench 2026-11-03 (BE9441): while monitoring, a MiniMate answers
the 0x2C call-home read with its DEVICE-INFO block ('Instantel', 'MiniMate Plus',
its serial) -- as it answers 1A -- and idle again, minutes later, the call-home
block. SFM writes back only a block it just read, so a monitoring MiniMate's
call-home cannot be changed; the change core already refused (byte 5 read as
'unknown(0x6E)', nothing sent), but said so cryptically.
- minimateplus.protocol.is_call_home_block(raw): the block starts with its own
length 0x7C and is N (or N + 2) bytes unescaped; the device-info block fails it.
- The Series-3 change raises UnitStateUnknown (409) with 'did not return its
call-home settings -- a MiniMate answers the call-home read with its
device-info block while monitoring; stop monitoring first ... nothing sent',
for a dry run too. call_home_state returns enabled = None with that reason.
- The page's callHomeRows shows the reason in place of 'Unknown (byte 5 = ...)'.
A blind write while monitoring stays untested on purpose.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
device_routing.js (Node-tested): call_home_state / call_home_on / call_home_off
routes (serial required); controlsFor gains callHomeChange (by serial, a
confirmed Series 3 or 4) and callHomeRead now covers a Series 4 by serial;
serialPanes gains callHomeForm / callHomeBox; callHomeRows renders the switch
(On / Off / 'Unknown (byte 5 = 0xNN)'), dial string and series;
changeErrorText takes the whole label ('Start monitoring', 'Enable call-home').
sfm_webapp.js: by serial the Call Home tab hides the legacy multi-field form
(Series 3 by host only) and shows the unit's switch with On / Off. Each asks
confirm(), sends one audited change, reports it with changeSummary /
changeErrorText, then re-reads quietly so the change's own result stays on
screen. A different unit's switch is never left showing. The new buttons are
wired with addEventListener, not onclick= (no inline JS).
Checked in jsdom against the real page: placeholder, read, Off (confirm, the
disable route, 'confirmed -- enabled -> disabled'), a serial edit clearing it,
legacy mode unchanged, no script errors.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- #6 Config tab by serial: the empty legacy edit form stays hidden (settings
are read-only until phase 5); the settings box shows from the start, 'Not read
yet', and a different unit's settings are never left on screen.
- #10 A start/stop that went out but was not confirmed is no longer called
'failed'. The route now adds 'sent' (acknowledged / unknown) and 'audit_id'
to a ChangeUnconfirmed -- same status, same message -- and the page says
'Start monitoring not confirmed: <what happened, audit record>'.
- #12 Device tab by serial: the empty legacy Compliance/Project tables are
hidden behind a pointer to the Config tab; a Micromate's Firmware card reads
'Thor line (version not read)' (its version string is in 0x01, undecoded).
- #16 A 504 with nothing sent adds: the connection opened, so the modem is
reachable -- another program (THOR or Blastware) may be holding the line, or
the unit is busy. Measured 2026-10-29: THOR polling leaves SFM a 0-byte 504.
New logic lives in sfm/static/device_routing.js (firmwareLabel,
deviceErrorText, changeErrorText, serialPanes) under Node tests; the page glue
in sfm_webapp.js only applies them. Checked in jsdom against the real page
(served at a fake origin, /static from disk): all four, legacy mode unchanged,
no script errors.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Stack-wide rule since 2026-11-02 (terra-view docs/tmi-stack.md, Terra-View's
rule 4): no inline JavaScript, because a test cannot load it. This page's tests
had to regex their code out of the HTML. Batch 2 edits both of the page's
scripts, so they come out first, verbatim:
- sfm/static/device_routing.js: the tested routing rules (157 lines)
- sfm/static/sfm_webapp.js: the rest of the page (2,274 lines)
The page loads them with <script src="static/..."> where the inline scripts
were, so load order and the onclick handlers are unchanged. Verified: putting
the files back inside those tags reproduces the old sfm_webapp.html byte for
byte, and a jsdom page check (served at http://sfm.local/, /static from disk,
no network) gives identical output before and after.
The server serves sfm/static/ at /static/, with Cache-Control: no-cache so a
browser revalidates (a cheap 304) and a new SFM never runs a cached old
script. Opened from disk, the page needs its static/ folder beside it.
Tests: the routing tests load the file instead of regexing the HTML; new
test_webapp_static pins no inline <script> in the page and the /static route
(through the whole app, middleware included).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- The arrow keys reached the legacy per-index waveform route in serial mode,
even with no list loaded (an unlocked MiniMate chain walk + 5A on the same
modem session). perIndexLive() is now the one rule for the per-index
buttons, the chips and the keyboard: live only for a list loaded without a
serial that has events.
- A failed reconnect left the previous unit's confirmation live (badge,
Diagnostics incl. the unguarded blind/slow-drip stops, call-home read).
connectUnit() now drops it before it identifies; only this answer confirms.
- Unit strings reached innerHTML: the Device tab's cards (active_setup, new in
1c) and its compliance/project tables now render as text, and Diagnostics'
error spans go through escapeHtml().
- CLAUDE.md: the ring-path Runs A/B are a merge gate (the spec's prerequisite
to a webapp sending serial), not only a use-time one.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- Config, with a serial: Read shows the unit's shared UnitSettings (setup,
record mode, sample rate, record time, histogram interval, geo trigger,
notes; — for null) in place of the write form. settingsRows() is in the
device-routing block, Node-tested; the rows are rendered as text, since
setup names and notes come from the unit.
- Write stays off for both series (and without a serial), tooltip
"Coming in phase 5 (#46)"; Call Home write likewise, "Coming in phase 4".
- Call Home read follows controlsFor().callHomeRead; for a confirmed Series 4
the tab says "Not yet available for Series 4".
- Diagnostics: every button follows controlsFor().diagnostics (Task 3), and a
refresh for anything but a confirmed Series-3 unit sends nothing and says
why. Its requests carry the typed serial, so the server guard checks it.
- Without a serial the legacy reads are unchanged.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- Load events with a serial lists the shared EventRef shape as a table: index,
time, type, PVS and the three peaks (or —), each row with its own Download.
- Download POSTs the row's event_id; SFM checks and stores it. The status bar
says what was stored (or that it already was), and the stored event opens
in the viewer. A refusal shows the server's detail verbatim.
- eventView() (device-routing block, Node-tested) reads both event shapes for
the table, the chips and the peaks bar.
- A list loaded without a serial keeps the chips and the per-index Load /
Save / Download buttons, unchanged; for a serial-loaded list those buttons
are off with a tooltip pointing at the rows.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- A Serial field beside host/port. With one, the Live tabs use the serial-based
endpoints (both series); without, every read behaves as before.
- The series badge, Hang up (Series 4), Diagnostics (Series 3), call-home read and
the disabled config/call-home writes all follow applyControls() <- controlsFor()
on the unit's own /device/info answer; editing the serial, host or port drops
the confirmation. Erase additionally needs a confirmed Series 3.
- Every /device/ request goes through deviceRequest/deviceFetch <- deviceUrl
(a static test enforces it); the legacy per-event routes joined deviceUrl
(event_waveform / event_file, index in the path).
- The status panel and Device tab read both shapes (shared and legacy).
- In serial mode the connect skips the legacy storage_range read (one connection
fewer; a Micromate has no such route). Start/stop without a serial behave as
before; with one, only for a confirmed series.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
One block of sfm_webapp.html (/* <device-routing> */) holds every rule that
decides where a /device/ request goes or whether a series-specific control is
live: seriesFromSerial, unitQuery, deviceUrl, confirmedSeries, controlsFor,
changeSummary, downloadSummary. No DOM, no fetch, so the tests run it in Node.
The safety rules it pins: a control follows the unit's own /device/info answer,
never a typed serial alone, and stops counting when the serial is edited;
Diagnostics only for a confirmed Series 3, Hang up only for Series 4; config and
call-home writes off for both series until phases 4-5. Also a test that the
page's inline scripts parse together (shared global scope).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44