SFM converted mic counts to psi with a '~0.5 psi full-scale' guess
(0.0125 / 32768 psi per count). Measured against 1,729,253 DL2
histogram intervals' 'MicL Peak' columns: one count is exactly 0.25 Pa.
count x 0.25 Pa matches Blastware's psi(L) on every interval, and
20 log10(count x 0.25 Pa / 20 uPa) its dB(L) on all but 17 exact
half-way cases.
- event_file_io: _MIC_FS_PSI -> _MIC_PSI_PER_COUNT = 0.25 / 6894.757
(3.626e-5 psi per count, was 3.81e-7). The event's mic peak in psi,
and with it the .h5 mic trace (event_hdf5 back-solves its scale from
that peak), come right on events with no report.
- waveform_codec.mic_count_to_db: the offset is 20 log10(0.25 / 20e-6)
= 81.938, not 81.94 -- the old value missed 49 of those intervals at a
rounding edge (count 119 printed 123.4, 81.94 gave 123.451).
The decoded count was always right: count 16 on K441LTON.900 is
Blastware's 106.0 dB.