SFM: Series-4 parity and a device safety layer — design
Date: 2026-10-20 · Branch: feat/series4-sfm-parity · Status: draft for review
Decided with: Brian, in conversation 2026-10-19/20.
In plain terms
SFM can already read a Micromate almost as well as it reads a MiniMate, but only through the Python library and bench scripts. SFM's web endpoints still only speak Series 3. This design does three things:
- Gives Series 4 the same everyday jobs in SFM that Series 3 has: identify a unit, check its status, read its settings, list and download events, start/stop monitoring, and change its settings.
- Adds a safety layer every device request goes through. It makes sure a command only reaches the unit it was meant for, logs every change, and checks every change by reading the unit back.
- Lets SFM stay on the line with a Micromate. It calls up once, confirms the serial, and keeps the connection open for later commands instead of reconnecting each time.
The guiding principle (Brian): talking to a Series 3 unit and talking to a
Series 4 unit is the same for the user. It is the same screens, the same
endpoints, the same way of naming a unit, and the same shape of answer. The
only visible difference is the serial (UM… vs BE…) and the settings each
model has. The differences live underneath, in one adapter per series.
It does not cover the emergency tools (runaway-unit stops, rescue, erase), the call-home receiver, or anything async. Those come later, as features shared by both series.
Production reach, stated honestly: while THOR is set to stay connected to production Micromates, a modem's single connection is taken. SFM can't reach those units until THOR lets go or SFM takes over call-home. In this phase, Series-4 SFM is for the bench and for units THOR is not holding.
Decisions
| decision | why | who |
|---|---|---|
| One user experience for both series | the user's method is identical; only the serial prefix and each model's settings differ; adapters absorb the rest | Brian |
| Match jobs, not endpoints | some Series-3 endpoints only exist to work around Series-3 quirks | agreed |
| No Series-4 emergency tools yet (runaway stop, rescue, erase) | Micromates rarely need them | Brian |
The series comes from the serial prefix: UM = Micromate, BE = MiniMate Plus, BA = Blastmate |
the prefixes are definitive, so there is no probing | Brian |
| Before any write, confirm the unit that answered is the one named | SFM addresses units by IP; a moved modem or stale IP means a different unit answers | agreed |
| Synchronous endpoints for now; async later, for both series together | parity first, then shared features | agreed |
| No library of setup files on the PC (unlike THOR). The unit's own settings are the truth: SFM reads them from the unit, and verifies every push by reading the unit back | THOR keeps one copy of each setup on the PC and another on the unit, linked only by file name. Nothing checks the contents match, so the PC's test2 and the unit's test2 can say different things |
Brian |
| Sessions for Series 4 now; Series 3 only after a bench test | THOR holds Micromate lines daily without harm; a MiniMate's hangup/DCD behaviour is untested | Brian |
| Audit log in SFM's database (new table) | Terra-View can show it; it fits the existing CREATE TABLE IF NOT EXISTS pattern |
proposed |
Scope: the parity table
Rows 8 and 11 of the original inventory (runaway stops, rescue/erase) are out of scope by decision. Row 6 (monitor log) moved to the parking lot: it is its own piece of work, together with call-in monitor-log ingest.
| # | job | Series 3 in SFM today | Series 4: protocol status | Series 4 in this phase |
|---|---|---|---|---|
| 1 | Identify a unit | /device/info (walks the whole chain, slow) |
implemented: connect() |
dispatch to the Micromate client |
| 2 | Quick status (monitoring, battery, memory, clock) | monitor/status, storage_range, events/index |
implemented: get_state(); the 0x06 event count is decoded only in the bench tool |
same endpoints; move the count decode into the client |
| 3 | Read settings | compliance in /device/info; GET /device/call_home |
get_setup() implemented and verified on hardware; call-home: enable + dial string observed, rest unmapped |
setup in full; call-home returns enable + dial string + raw |
| 4 | List events | /device/events (downloads every event to list them) |
implemented, and cheap: the chain walk returns sizes | the same event-list shape as Series 3, each event carrying an event_id |
| 5 | Download an event | /event/{i}, /waveform, /blastware_file |
implemented: streams the .IDFW/.IDFH itself, with a decode self-check |
the same download and save-to-DB actions as Series 3. Underneath, Series 4 saves through /db/import/idf_file |
| 7 | Start / stop monitoring | monitor/start, monitor/stop |
mapped, identical to Series 3 (0x96/0x97, acks 0x69/0x68); never sent |
first write. Verify with 0x49 state |
| 9 | Change call-home settings | POST /device/call_home |
write mapped (0x2C → 0x7E/0x7F, same as Series 3); field map mostly unknown |
enable/disable only; the rest waits for the THOR round |
| 10 | Change recording settings | POST /device/config |
sequence mapped (0xDA → 0x68/73 → 0x82/83 → 0x71/72); no encoder |
read-modify-write, stop → push → restart, verify with get_setup() |
Architecture
One door for every device request
A new module, sfm/device_access.py, sits between the /device/* handlers and
the two device libraries. No handler opens a transport or picks a library
itself any more.
/device/* handler
│ (serial, address, action, tier)
▼
device_access ── series from serial prefix ── session / per-unit lock
│ │ │
│ ▼ ▼
│ minimateplus (S3) or micromate (S4) audit table
▼
read-back verification (changes only)
- Naming a unit is the same for both:
serial+ address. The prefix picks the adapter:UM→ Series 4 (micromate);BEorBA→ Series 3 (minimateplus); any other prefix → 422. A request with noserialis treated as Series 3. That is a legacy path kept only so existing scripts keep working; the webapp and every new caller always sendserial. - One contract. Same
/device/*paths, same request parameters, and the same response models for both series:UnitInfo,UnitStatus,UnitSettings,EventRef,ChangeResult- Anything only one series has goes under a
detailobject inside the same model, never as a different top-level shape."series": 3 | 4is informational.
- Events have one identity. An opaque
event_idstring from the event list, passed back to download it. Underneath it is Series 3's position or Series 4's device key, plus the serial; the user never handles either. - Not-yet-supported is uniform. If one series can't do something yet, the endpoint still exists and answers 501 "not supported for Series N yet". It is never a missing route or a different call.
- Work separate from HTTP. Each operation is a plain function taking the session. The handler is a thin wrapper, so making an operation an async job later means changing the wrapper, not the operation.
Adapters
Each series implements the same operation set behind device_access:
| operation | Series 3 adapter | Series 4 adapter |
|---|---|---|
identify |
STARTUP + serial + config | connect() |
status |
0x1C (+ 0x06) |
get_state() + 0x06 count |
read_settings |
0x1A compliance decode |
get_setup() |
list_events |
today's walk | chain walk (cheap) |
download_event |
5A → Blastware file | 5A → .IDFW/.IDFH |
start / stop |
0x96 / 0x97 (exists) |
0x96 / 0x97 (new) |
read_call_home / set_call_home_enabled |
exists | 0x2C / 0x7E+0x7F |
push_settings |
/device/config path (exists) |
setup push (new) |
call_up / hang_up |
connect-per-request underneath, until the bench test | held session |
Each adapter maps its native data into the shared models. UnitSettings keeps
the fields both series have under the same names and units: record mode,
sample rate, record time, histogram interval, trigger levels, the notes as
label → value. Model-specific fields go under detail.
Requests that include serial get the shared models; requests without it
keep today's Series-3 shapes, unchanged. Nothing breaks in between: the
webapp moves to serial in phase 1c, and the old path retires after that and
after watch_unit.sh moves. Terra-View does not call /device/*. The
emergency endpoints (stop_monitoring_*, rescue, events/erase) are not
touched and stay Series-3 only; a UM serial gets 501.
The webapp
The same screens for both series. You name a unit by serial + address (today it takes host/port only), and the series shows as a badge, nothing more. Read panels for both series land in phase 1; each write button appears for Series 4 in the phase that adds that write.
The safety layer
Tiers. Every operation declares one:
| tier | examples | what the layer does |
|---|---|---|
| read | info, status, setup, events, download | runs; no identity check |
| change | start/stop, call-home, setup push | identity check → audit → send → read back → audit the result |
| destructive | delete, erase (none in this phase) | everything above + the caller must send the unit's serial as confirm |
Identity check (changes only). Before the first change in a connection,
compare the serial the unit reports with the one requested; on a mismatch,
refuse (409) and send nothing. Series 4 already reads the serial in
connect(). A Series-3 change adds one serial read (SUB 0x15), about one
round trip.
Opcode guard. A send-side check that refuses known cross-series
collisions. It starts with 0xA2, which is CONFIRM ERASE on Series 3 and
"end call-home" on Series 4. This phase never sends 0xA2; the guard makes
that impossible by accident. The table of colliding opcodes lives in one
place.
One connection per unit. A per-unit lock keyed by address
(tcp:host:port / serial:port). A request for a busy unit waits up to
20 s (configurable; under Terra-View's 60 s proxy timeout), then gets
423 "unit busy" instead of opening a second connection. This applies to Series 3 as well. The webapp's Diagnostics
refresh currently fires three device requests at once against a modem that
allows one; under the lock they queue.
Read-back verification. A zero-data ack has never been proof that a write applied. Every change is followed by a read:
- start/stop →
0x49state: start expects0x02, stop expects0x00. Record the value either way;0x03is the in-call-home state - call-home → re-read
0x2Cand compare the edited bytes - setup push →
get_setup()and compare every field that was set
A mismatch is reported as a failed change (and audited as one), even if the ack was clean.
Errors that say what happened. Distinct responses for:
- 503: couldn't connect
- 504: connected but the unit didn't answer
- 423: unit busy
- 409: wrong unit
- 422: bad input
- 502: the reply was malformed
Today 502 covers both of the first two.
Audit table. New table device_commands, created on startup like the
others. It is additive, but still a schema change, so the changelog says so.
One row per change attempt:
| column | content |
|---|---|
ts, serial, series, address |
who and where |
action, tier |
e.g. monitor_stop, change |
request |
parameters, plus hex of what was sent |
ack |
reply summary |
verified |
read-back result: ok / mismatch / not_checked |
outcome, error |
final result |
caller |
best-effort origin (webapp, terra-view, script) |
Reads are not audited.
Sessions
- Call up. Connect,
connect()(POLL / SERIAL / STATE), identity check against the requested serial. - On the line. Requests for that unit run in turn on the same connection, behind the per-unit lock.
- Keep-alive. A POLL every 60 s while idle. The ACEmanager TCP idle timeout is 2 min.
- Hang up. On
POST /device/hangup, after 5 idle minutes, or on any transport error. A reconnect is a new session with a new identity check. - Ring safety. If
RADIO RINGarrives on an SFM session, log it and hang up immediately. Until SFM is the call-home destination, the unit's retry should reach THOR. 🔶 Inferred: the unit retries; its0x2Cretries field reads 3.
Call up and hang up are the same for both series. Underneath, the Series-3 adapter keeps today's connect-per-request behaviour until the bench test under Testing; the user sees the same actions either way.
Series-4 write path
Encoder. Extend micromate/framing.py:
- a request builder that carries a data section
- the same escape set
{02,03,04,10}, applied to data, params and checksum - plain SUM8 of the destuffed payload (not Series 3's write format)
Rules: micromate_protocol_reference.md → The escape set — exactly four bytes,
Checksum — plain SUM8 of the de-stuffed payload, Write-frame destuffing.
Dry run. Every Series-4 write accepts dry_run=true. It returns the exact
frames it would send and sends nothing. The first live send of each write is
reviewed this way first.
Per operation:
| operation | frames | verify | open questions it settles |
|---|---|---|---|
| start / stop | 0x96 / 0x97, offset 0xFFFF, zero params |
0x49 |
the ack when already in that state |
| call-home enable/disable | read 0x2C → 0x7E (126 B content, offset 0x007E) → 0x7F |
re-read 0x2C |
none new; bit 0 of write[5] |
| setup push | read 0x08, 0x2E, 0x1A → 0xDA (name) → 0x68/0x73 → 0x82/0x83 → 0x71/0x72 |
get_setup() |
the offset rule; whether 0x68/0x82 are needed |
Setup push, details:
- Read-modify-write of all three blocks, as THOR appears to do.
🔶 Inferred from two captures:
0x68= the preferences content with[0:3]zeroed;0x82= the0x2Econtent with[0:3]and the CRC slot zeroed;0x71= the0x1Acontent[0:2090]with the length echo zeroed. - Offsets come from THOR's observed constants. 🔶 The hypothesis "offset = the read block's length + 2" fits all four, but each has been seen at one size only.
- The edited fields come from
MicromateSetupand are written back at the offsetssetup_block.pyalready uses. Only fields graded measured may be edited. - Sequence: stop monitoring (if running) → push → restart (if it was running) → verify. THOR's "must be idle" is policy, not the device, but stopping first removes a variable from the first live write.
- ⚠ Edit-buffer hazard.
0x1Areads the keypad's unsaved edits, so a read-modify-write taken while someone is in the menu would commit them. This can't be detected from the wire; it is documented in the endpoint and audited. - Target file: the active setup, overwritten in place, by default. A
named target is allowed, but overwriting a non-active setup or
factory.MMBis untested, so it is refused in this phase.
Testing
Offline (CI-capable):
- Frame reproduction. The new encoder must reproduce THOR's captured write
frames byte for byte:
0x96/0x97,0x7E/0x7F, and the full setup push, from the 9-24 / 9-25 / 10-14 captures. This is the same method that proved the reads (218/218). The needed frames are copied into force-added fixtures, sincebridges/captures/is gitignored. - Safety layer on fakes:
- series by prefix
- identity mismatch → 409, nothing sent
0xA2refused- lock → 423
- read-back mismatch → failed change
- audit row written for every change attempt
- RADIO RING → session ends
- Endpoints through FastAPI's
TestClientagainst the scripted transport already used intests/test_micromate_client.py. - Contract tests. The same test, run against both adapters with fakes,
asserts that each job returns the same model with the same field names and
units. A field present for one series and absent for the other is a test
failure unless it lives under
detail. This is the test that keeps the guiding principle true as the code grows.
Bench (UM20147, with Brian): each write's first live send follows the same protocol:
- Dry run.
- Review the frames.
- Send once, with a capture running.
- Read back.
- Write the result into the reference.
Order: stop/start → call-home off/on → setup push of one field → setup push of several fields.
Series-3 session test (before Series-3 sessions are built): hold a call-up session open to a MiniMate that is monitoring, for several minutes. Record:
- does it keep monitoring (does an event still record)?
- after hangup, is it monitoring?
- does it behave differently on an RV55 with the DCD issue?
Build phases
Each safety piece arrives with its first real use, so nothing sits untested.
- One contract, reads only.
device_access.py, the two adapters, the shared models, series selection, the per-unit lock, the error mapping, and sessions with ring safety. Rows 1–5 for both series through the same endpoints. No writes. Split into three plans:- 1a. The foundation plus identify / status / settings for both series.
- 1b. Event list and download for both series (rows 4–5). Decided with
Brian 2026-10-21:
- MiniMate events are listed from their headers (
0A+1Earm +0C+1Fbrowse, no5A) instead of downloading every event to list it, and one event downloads without the ones before it. It reuses the bypassskip_waveform_for_eventsalready takes, so no new command reaches the wire. The legacy serial-less/device/eventskeeps its full-download walk. 🔶 Unproven on a unit: a MiniMate bench check is owed before this branch merges. Its scope (final review, 2026-10-21):- a list of three or more events, confirming
1F(browse)advances after a1E(arm)without1F(arm); - a non-first event;
- a histogram-mode event, capturing its
0C(which would let Series 3'srecord_type,nulltoday, be decoded); - an event past 64 KB, expecting the named 502;
- a stale-id download, confirming no
5Agoes out; - a download of an event Blastware's call-home already delivered (one
waveform, one histogram), expecting
already_stored. This is the only way to measure that Series 3's (serial, timestamp) agrees between the unit's0Cand the watcher's import; if it does not, a download silently stores a duplicate row (never a downgrade). Measured for Series 4: 1,016/1,016 watcher imports found by the lookup (2026-10-21). - a dropped reply, via
bridges/mm_link.py-style fault injection.
- a list of three or more events, confirming
- Download = fetch, check, store in SFM's database, the same as
Series 3's
blastware_fileroute, and returns the stored row's id. The file and waveform are then served by the existing/db/events/{id}/…routes. The database matches on (serial, timestamp), so storing an event already stored, including one THOR's watcher forwarded,refreshes the row instead of duplicating it. Changed by the final review (2026-10-21): it returns the existing row and writes nothing (inserted: false,detail.already_stored: true). Never merge: the watcher's copy, with its paired report, is the richer one, and a refresh measurably downgraded it (PVS, project strings and ZC frequencies went NULL). event_id=<serial>:<key>:<YYYYmmddHHMMSS>. Both series reuse keys (after an erase, and across Micromates), so a download whose key now holds a different event fails with 409 instead of fetching it.- Micromate downloads walk to the event first (the only order captured from THOR), with THOR's 1,024-byte chunks.
- MiniMate events are listed from their headers (
- 1c. The webapp: serial + address, and the same read panels for both
series.
- HARD PREREQUISITE: before any poller or webapp sends
serial, add a per-address cool-down after a RADIO RING: call-ups to that address answer 503 +Retry-Afteruntil it expires, andPOST /device/hangupclears it. The length must come from the unit's measured call-home retry timing (a bench test). Without it a poller can re-grab the modem's single TCP session through every retry, and the call never reaches THOR.
- HARD PREREQUISITE: before any poller or webapp sends
- Encoder + opcode guard + dry run + frame-reproduction tests. Still
nothing sent.
(2026-10-22) "Dry run" in this phase means the write plan: the exact
frames, in order, with their expected acks, built from the unit's own reads
(
micromate.writes), plus a bench tool that prints one from a live unit's reads (bridges/mm_dry_run.py). Each write endpoint'sdry_run=truearrives with that write in phases 3-5 and returns the plan. The opcode guard is a send-side gate inMicromateProtocol._send: every write SUB is refused unless the connection was built with it allowed (none is, in this phase), and0xA2is refused always. - Start/stop monitoring (row 7): the first live write, at the bench. Brings the identity check, the audit table and the read-back framework.
- Call-home enable/disable (row 9).
- Setup push (row 10): one field first, then several.
Each phase is its own implementation plan. All phases land on
feat/series4-sfm-parity and merge to dev once, when everything here is
implemented (Brian, 2026-10-21) — not one merge per phase.
Out of scope (this phase)
- Series-4 emergency tools: runaway stops, rescue, per-event delete, erase. (Erase-all is also unmapped.)
- Async jobs; auth (Terra-View as jump host); the call-home receiver. These are shared features for later.
- Scheduler read/write/enable; clock set; remote sensor check; restart; firmware.
- Series-3 sessions, pending the bench test.
- Mapping the rest of
0x2C(the THOR round).
Fix list (found 2026-10-19, separate small changes)
bridges/ach_server.py:524:seen_keysis undefined. Series-3 call-home monitor-log collection has failed every session since9afa348(2026-05-07).bridges/mm_ach_server.py:262:_cstring/_contentaren't imported, so the ring-answer path would raiseNameError. From8f67608(2026-10-14).- Series-3
5A64 KB page-boundary bug (known; surfaces as HTTP 500). POST /device/configwrites a fixed 29-byte0x82blob from one 3-11-26 capture, whatever the unit's trigger settings./device/eventsruns a full 5A download per event just to list headers.0x08is namedread_event_indexinmicromate/; it is the preferences block.
Left open by the phase 1a review (2026-10-21)
Minor findings from the whole-branch review, parked here so they outlive the plan's scratch ledger. None blocks 1a; the phase each one matters to is named.
- CORS ordering (1c, if a browser calls SFM directly). The
/deviceguard middleware is registered after CORS, so it sits outside it: its 409/501 responses carry noaccess-control-allow-origin, and an OPTIONS preflight to a guarded URL gets the guard's refusal instead of CORS's 200. Register CORS last, or let OPTIONS through. Terra-View proxies server-side and is not affected. - Legacy handlers take no unit lock (1c / phase 3). The guard checks for a held line and then lets the legacy handler run unlocked; a legacy request admitted just before a Series-4 call-up at the same address can overlap it.
- Address keys are not normalised.
tcp:<host>:<port>is the raw string: a hostname vs an IP, or/dev/serial/by-id/...vs/dev/ttyACM0, gets two locks for one link. - A serial-less request to an UNHELD Micromate still runs Series-3 bytes. The guard only knows about lines SFM holds. This is the phase-2 opcode guard's job.
tcp_portcoercion mismatch. The guard parses withint(), pydantic coerces more loosely ("9034.0"), so such a request passes the guard unchecked. Parse with pydantic's own int coercion, or fail closed.- Test gaps. The middleware wrapper itself is untested (a raw ASGI call
works without httpx); the contract test compares field names of one shared
dataclass, which cannot fail, and does not enforce "a field absent for one
series lives under
detail". _ensure_reaperrace. Two first Series-4 requests at once can start two reaper threads; harmless becausetry_holdkeeps them off each other's units.- Held serial lines share
/dev/ttyACM0silently. pyserial's open is not exclusive, so a bench script beside a held line interleaves bytes with the keep-alive; it looks like a dead unit. Open withexclusive=True, or don't hold serial links.
Left open by the phase 1b review (2026-10-21)
Parked from the whole-branch review of 1b; none blocks it. The phase each matters to is named.
- Retry re-runs a whole download (1c). The retry-once on a held-line timeout re-runs the full download, roughly doubling time-to-error; narrow it to failures before the first reply.
last_usedafter a kept-line error (1c). It is not refreshed when an op's verdict (404/409) keeps the line, so the idle hang-up comes early.verified: nullis ambiguous (1c). It means both "a histogram" (no check exists) and "a waveform whose check could not run".- Mic naming and the Series-4 mic decode (1c).
detail.mic_psion Series 3,detail.micon Series 4; Series 4's0x0Cmic decodes as a denormal (~5e-39) — fix the decode before 1c shows it. - 500 detail leaks paths (1c, before SFM is reachable beyond Terra-View). The download handler's 500 detail includes raw exception text, filesystem paths included.
- Serial as a directory name (1c). The serial is never checked against
[A-Z0-9]+before it names a store directory; only the unit can choose it, but the check is cheap. - No server-level RADIO RING mid-download test (1c). Covered structurally (DeviceAccess-level test + the ring transport), not through the handler.
- Series-3 stored
record_type(MiniMate bench check). Rows SFM stores for Series 3 still say "Waveform" for histograms — pre-existing, the same as the legacyblastware_fileroute; fix it with the0Cdecode. - A fresh Series-4 download's gaps (1c). No
sample_rateis stored, and the sidecar has no vector sum; the DB row has the unit's PVS (overlaid from0x0C). - Orphan files (SDM). An event file in the store with no DB row is still overwritten by a download.
- Same-second collisions (SDM). Two events of one unit in the same second collide on (serial, timestamp) — a DB-wide dedup property, not 1b's.
- 423 without
Retry-After(1c). AddRetry-Afterto the 423 a request gets while a download holds the unit. - A lost
0Cduring a Series-3 list (1c). The strict walk raises on a lost0Aor1F, but a lost0Cstill lists the event with no timestamp and anevent_idending:-. Downloading that id skips the stale check and runs the full5A. If the0Cis lost again it is refused (no timestamp, no filename); if it reads, whatever event now sits at that key is stored with no stale check: self-consistent data, but possibly not the event that was listed. Fix: raise on a lost0Cin strict mode, or refuse:-ids for Series 3. - A lost reply on a Series-3 TCP walk re-runs it once (1c) under the adapter's cold-boot retry before the 504, doubling time-to-error on a bad link.
- A Series-3 download of an already-stored event still runs the full
5Abefore the database lookup finds it (the lookup is on the SDM side, after the device layer). Wasted transfer, not bad data (1c).
Table 2: to be added later (both series)
| area | item |
|---|---|
| infrastructure | async jobs; auth via Terra-View |
| THOR has it, SFM doesn't | clock set; remote sensor check; scheduler; one setup to many units; restart |
| call-home receiver | hold lines open (S4); "already downloaded" tracking; rules by serial prefix (copy events, monitor log, delete, set clock); detect "Monitor While Calling Home" blocking deletes (⚠ one unexplained 21 → 5 count drop in THOR's log); hand off to /db/import/* instead of the watchers |
| settings, the no-PC-files way | intended settings per deployment; drift report (intended vs actual); push then verify |
| notifications | email/SMS by event type and PPV (THOR does it; may belong in Terra-View) |
| emergency tools for S4 | if a Micromate ever needs them |
Parking lot
Monitor logs. Its own piece of work, data side as much as device side.
A unit calls in at least once or twice a day. With no events (e.g. waveform
mode, no blast that day) it still delivers a monitor log: "monitoring from
X to Y". The watchers (one for BW-ACH / Series 3, one for THOR-ACH / Series 4)
use the newest file per unit, event or .MLG, as SFM's "last heard from" age.
They forward waveform and histogram files but not monitor logs. To decide:
- Ingest call-in monitor logs into SFM. SFM's DB already has a
monitor_logtable (one row per interval, deduped by serial + start time), andminimateplus/blastware_file.pyalready parses Series-3.MLG. Missing: anything that takes.MLGin from the watchers, and knowledge of THOR's Series-4 monitor-log file format. - A live monitor-log read over a connection (Series 4:
0x0Aexists, no wrapper or decoder; Series 3 in SFM: none). - What it is for: proof of coverage ("was it monitoring when they blasted?") and visible gaps.
- Related fix-list item:
ach_server.py'sseen_keys.
How SFM handles setup files and schedules. This phase only reads and
pushes the unit's active setup, with no PC-side library (the .mmb
same-name/different-contents gotcha). Still to decide, as one question because
the two are coupled:
- The unit stores up to 25 named
.MMBfiles. Should SFM manage them (list, select, rename, delete, push to a named file), or only ever touch the active one? - Schedules name setup files. A schedule entry switches the unit to a setup by name, so a schedule is only as reliable as the files it points to. What if it points to a file that isn't there? That is untested.
- How "intended settings per deployment" (Table 2) fits with files and schedules.
- Scheduler read/write is mostly mapped (file transfer +
0x47); disable has never been captured.
A modem-side agent on the RX55. AirLink
OS on the RX55 has "Container Applications (Early Adopter)", and Brian already
runs slmm-proxy there. A small program owning the unit's serial port could:
- let THOR and SFM share the line
- dial out to SFM, avoiding the IP whitelist
- handle
RADIO RINGlocally - Brian's hunch: make RX55s work with Series 3 by driving DTR/DCD itself
Open questions:
- Can a container open the RS-232 port?
- Can it set the control lines?
- How is it deployed to a fleet, and does it Autostart?
- Will a firmware update keep the beta feature?
Open questions this design leaves for the bench
- The ack for start/stop when the unit is already in that state.
- The offset rule, and whether
0x68/0x82are required, for the setup push. - Whether the device ever refuses a write while monitoring. Only THOR's policy has been seen.
- How a failed write is signalled. It has never been observed.
- MiniMate behaviour with a held session (above).