SFM: Series-4 parity and a device safety layer — design

Date: 2026-10-20 · Branch: feat/series4-sfm-parity · Status: draft for review Decided with: Brian, in conversation 2026-10-19/20.

In plain terms

SFM can already read a Micromate almost as well as it reads a MiniMate, but only through the Python library and bench scripts. SFM's web endpoints still only speak Series 3. This design does three things:

  1. Gives Series 4 the same everyday jobs in SFM that Series 3 has: identify a unit, check its status, read its settings, list and download events, start/stop monitoring, and change its settings.
  2. Adds a safety layer every device request goes through. It makes sure a command only reaches the unit it was meant for, logs every change, and checks every change by reading the unit back.
  3. Lets SFM stay on the line with a Micromate. It calls up once, confirms the serial, and keeps the connection open for later commands instead of reconnecting each time.

The guiding principle (Brian): talking to a Series 3 unit and talking to a Series 4 unit is the same for the user. It is the same screens, the same endpoints, the same way of naming a unit, and the same shape of answer. The only visible difference is the serial (UM… vs BE…) and the settings each model has. The differences live underneath, in one adapter per series.

It does not cover the emergency tools (runaway-unit stops, rescue, erase), the call-home receiver, or anything async. Those come later, as features shared by both series.

Production reach, stated honestly: while THOR is set to stay connected to production Micromates, a modem's single connection is taken. SFM can't reach those units until THOR lets go or SFM takes over call-home. In this phase, Series-4 SFM is for the bench and for units THOR is not holding.


Decisions

decision why who
One user experience for both series the user's method is identical; only the serial prefix and each model's settings differ; adapters absorb the rest Brian
Match jobs, not endpoints some Series-3 endpoints only exist to work around Series-3 quirks agreed
No Series-4 emergency tools yet (runaway stop, rescue, erase) Micromates rarely need them Brian
The series comes from the serial prefix: UM = Micromate, BE = MiniMate Plus, BA = Blastmate the prefixes are definitive, so there is no probing Brian
Before any write, confirm the unit that answered is the one named SFM addresses units by IP; a moved modem or stale IP means a different unit answers agreed
Synchronous endpoints for now; async later, for both series together parity first, then shared features agreed
No library of setup files on the PC (unlike THOR). The unit's own settings are the truth: SFM reads them from the unit, and verifies every push by reading the unit back THOR keeps one copy of each setup on the PC and another on the unit, linked only by file name. Nothing checks the contents match, so the PC's test2 and the unit's test2 can say different things Brian
Sessions for Series 4 now; Series 3 only after a bench test THOR holds Micromate lines daily without harm; a MiniMate's hangup/DCD behaviour is untested Brian
Audit log in SFM's database (new table) Terra-View can show it; it fits the existing CREATE TABLE IF NOT EXISTS pattern proposed

Scope: the parity table

Rows 8 and 11 of the original inventory (runaway stops, rescue/erase) are out of scope by decision. Row 6 (monitor log) moved to the parking lot: it is its own piece of work, together with call-in monitor-log ingest.

# job Series 3 in SFM today Series 4: protocol status Series 4 in this phase
1 Identify a unit /device/info (walks the whole chain, slow) implemented: connect() dispatch to the Micromate client
2 Quick status (monitoring, battery, memory, clock) monitor/status, storage_range, events/index implemented: get_state(); the 0x06 event count is decoded only in the bench tool same endpoints; move the count decode into the client
3 Read settings compliance in /device/info; GET /device/call_home get_setup() implemented and verified on hardware; call-home: enable + dial string observed, rest unmapped setup in full; call-home returns enable + dial string + raw
4 List events /device/events (downloads every event to list them) implemented, and cheap: the chain walk returns sizes the same event-list shape as Series 3, each event carrying an event_id
5 Download an event /event/{i}, /waveform, /blastware_file implemented: streams the .IDFW/.IDFH itself, with a decode self-check the same download and save-to-DB actions as Series 3. Underneath, Series 4 saves through /db/import/idf_file
7 Start / stop monitoring monitor/start, monitor/stop mapped, identical to Series 3 (0x96/0x97, acks 0x69/0x68); never sent first write. Verify with 0x49 state
9 Change call-home settings POST /device/call_home write mapped (0x2C → 0x7E/0x7F, same as Series 3); field map mostly unknown enable/disable only; the rest waits for the THOR round
10 Change recording settings POST /device/config sequence mapped (0xDA → 0x68/73 → 0x82/83 → 0x71/72); no encoder read-modify-write, stop → push → restart, verify with get_setup()

Architecture

One door for every device request

A new module, sfm/device_access.py, sits between the /device/* handlers and the two device libraries. No handler opens a transport or picks a library itself any more.

/device/* handler
      │  (serial, address, action, tier)
      ▼
device_access ── series from serial prefix ── session / per-unit lock
      │              │                               │
      │              ▼                               ▼
      │      minimateplus (S3)  or  micromate (S4)   audit table
      ▼
read-back verification (changes only)
  • Naming a unit is the same for both: serial + address. The prefix picks the adapter: UM → Series 4 (micromate); BE or BA → Series 3 (minimateplus); any other prefix → 422. A request with no serial is treated as Series 3. That is a legacy path kept only so existing scripts keep working; the webapp and every new caller always send serial.
  • One contract. Same /device/* paths, same request parameters, and the same response models for both series:
    • UnitInfo, UnitStatus, UnitSettings, EventRef, ChangeResult
    • Anything only one series has goes under a detail object inside the same model, never as a different top-level shape. "series": 3 | 4 is informational.
  • Events have one identity. An opaque event_id string from the event list, passed back to download it. Underneath it is Series 3's position or Series 4's device key, plus the serial; the user never handles either.
  • Not-yet-supported is uniform. If one series can't do something yet, the endpoint still exists and answers 501 "not supported for Series N yet". It is never a missing route or a different call.
  • Work separate from HTTP. Each operation is a plain function taking the session. The handler is a thin wrapper, so making an operation an async job later means changing the wrapper, not the operation.

Adapters

Each series implements the same operation set behind device_access:

operation Series 3 adapter Series 4 adapter
identify STARTUP + serial + config connect()
status 0x1C (+ 0x06) get_state() + 0x06 count
read_settings 0x1A compliance decode get_setup()
list_events today's walk chain walk (cheap)
download_event 5A → Blastware file 5A → .IDFW/.IDFH
start / stop 0x96 / 0x97 (exists) 0x96 / 0x97 (new)
read_call_home / set_call_home_enabled exists 0x2C / 0x7E+0x7F
push_settings /device/config path (exists) setup push (new)
call_up / hang_up connect-per-request underneath, until the bench test held session

Each adapter maps its native data into the shared models. UnitSettings keeps the fields both series have under the same names and units: record mode, sample rate, record time, histogram interval, trigger levels, the notes as label → value. Model-specific fields go under detail.

Requests that include serial get the shared models; requests without it keep today's Series-3 shapes, unchanged. Nothing breaks in between: the webapp moves to serial in phase 1c, and the old path retires after that and after watch_unit.sh moves. Terra-View does not call /device/*. The emergency endpoints (stop_monitoring_*, rescue, events/erase) are not touched and stay Series-3 only; a UM serial gets 501.

The webapp

The same screens for both series. You name a unit by serial + address (today it takes host/port only), and the series shows as a badge, nothing more. Read panels for both series land in phase 1; each write button appears for Series 4 in the phase that adds that write.

The safety layer

Tiers. Every operation declares one:

tier examples what the layer does
read info, status, setup, events, download runs; no identity check
change start/stop, call-home, setup push identity check → audit → send → read back → audit the result
destructive delete, erase (none in this phase) everything above + the caller must send the unit's serial as confirm

Identity check (changes only). Before the first change in a connection, compare the serial the unit reports with the one requested; on a mismatch, refuse (409) and send nothing. Series 4 already reads the serial in connect(). A Series-3 change adds one serial read (SUB 0x15), about one round trip.

Opcode guard. A send-side check that refuses known cross-series collisions. It starts with 0xA2, which is CONFIRM ERASE on Series 3 and "end call-home" on Series 4. This phase never sends 0xA2; the guard makes that impossible by accident. The table of colliding opcodes lives in one place.

One connection per unit. A per-unit lock keyed by address (tcp:host:port / serial:port). A request for a busy unit waits up to 20 s (configurable; under Terra-View's 60 s proxy timeout), then gets 423 "unit busy" instead of opening a second connection. This applies to Series 3 as well. The webapp's Diagnostics refresh currently fires three device requests at once against a modem that allows one; under the lock they queue.

Read-back verification. A zero-data ack has never been proof that a write applied. Every change is followed by a read:

  • start/stop → 0x49 state: start expects 0x02, stop expects 0x00. Record the value either way; 0x03 is the in-call-home state
  • call-home → re-read 0x2C and compare the edited bytes
  • setup push → get_setup() and compare every field that was set

A mismatch is reported as a failed change (and audited as one), even if the ack was clean.

Errors that say what happened. Distinct responses for:

  • 503: couldn't connect
  • 504: connected but the unit didn't answer
  • 423: unit busy
  • 409: wrong unit
  • 422: bad input
  • 502: the reply was malformed

Today 502 covers both of the first two.

Audit table. New table device_commands, created on startup like the others. It is additive, but still a schema change, so the changelog says so. One row per change attempt:

column content
ts, serial, series, address who and where
action, tier e.g. monitor_stop, change
request parameters, plus hex of what was sent
ack reply summary
verified read-back result: ok / mismatch / not_checked
outcome, error final result
caller best-effort origin (webapp, terra-view, script)

Reads are not audited.

Sessions

  1. Call up. Connect, connect() (POLL / SERIAL / STATE), identity check against the requested serial.
  2. On the line. Requests for that unit run in turn on the same connection, behind the per-unit lock.
  3. Keep-alive. A POLL every 60 s while idle. The ACEmanager TCP idle timeout is 2 min.
  4. Hang up. On POST /device/hangup, after 5 idle minutes, or on any transport error. A reconnect is a new session with a new identity check.
  5. Ring safety. If RADIO RING arrives on an SFM session, log it and hang up immediately. Until SFM is the call-home destination, the unit's retry should reach THOR. 🔶 Inferred: the unit retries; its 0x2C retries field reads 3.

Call up and hang up are the same for both series. Underneath, the Series-3 adapter keeps today's connect-per-request behaviour until the bench test under Testing; the user sees the same actions either way.

Series-4 write path

Encoder. Extend micromate/framing.py:

  • a request builder that carries a data section
  • the same escape set {02,03,04,10}, applied to data, params and checksum
  • plain SUM8 of the destuffed payload (not Series 3's write format)

Rules: micromate_protocol_reference.md → The escape set — exactly four bytes, Checksum — plain SUM8 of the de-stuffed payload, Write-frame destuffing.

Dry run. Every Series-4 write accepts dry_run=true. It returns the exact frames it would send and sends nothing. The first live send of each write is reviewed this way first.

Per operation:

operation frames verify open questions it settles
start / stop 0x96 / 0x97, offset 0xFFFF, zero params 0x49 the ack when already in that state
call-home enable/disable read 0x2C → 0x7E (126 B content, offset 0x007E) → 0x7F re-read 0x2C none new; bit 0 of write[5]
setup push read 0x08, 0x2E, 0x1A → 0xDA (name) → 0x68/0x73 → 0x82/0x83 → 0x71/0x72 get_setup() the offset rule; whether 0x68/0x82 are needed

Setup push, details:

  • Read-modify-write of all three blocks, as THOR appears to do. 🔶 Inferred from two captures: 0x68 = the preferences content with [0:3] zeroed; 0x82 = the 0x2E content with [0:3] and the CRC slot zeroed; 0x71 = the 0x1A content [0:2090] with the length echo zeroed.
  • Offsets come from THOR's observed constants. 🔶 The hypothesis "offset = the read block's length + 2" fits all four, but each has been seen at one size only.
  • The edited fields come from MicromateSetup and are written back at the offsets setup_block.py already uses. Only fields graded measured may be edited.
  • Sequence: stop monitoring (if running) → push → restart (if it was running) → verify. THOR's "must be idle" is policy, not the device, but stopping first removes a variable from the first live write.
  • ⚠ Edit-buffer hazard. 0x1A reads the keypad's unsaved edits, so a read-modify-write taken while someone is in the menu would commit them. This can't be detected from the wire; it is documented in the endpoint and audited.
  • Target file: the active setup, overwritten in place, by default. A named target is allowed, but overwriting a non-active setup or factory.MMB is untested, so it is refused in this phase.

Testing

Offline (CI-capable):

  • Frame reproduction. The new encoder must reproduce THOR's captured write frames byte for byte: 0x96/0x97, 0x7E/0x7F, and the full setup push, from the 9-24 / 9-25 / 10-14 captures. This is the same method that proved the reads (218/218). The needed frames are copied into force-added fixtures, since bridges/captures/ is gitignored.
  • Safety layer on fakes:
    • series by prefix
    • identity mismatch → 409, nothing sent
    • 0xA2 refused
    • lock → 423
    • read-back mismatch → failed change
    • audit row written for every change attempt
    • RADIO RING → session ends
  • Endpoints through FastAPI's TestClient against the scripted transport already used in tests/test_micromate_client.py.
  • Contract tests. The same test, run against both adapters with fakes, asserts that each job returns the same model with the same field names and units. A field present for one series and absent for the other is a test failure unless it lives under detail. This is the test that keeps the guiding principle true as the code grows.

Bench (UM20147, with Brian): each write's first live send follows the same protocol:

  1. Dry run.
  2. Review the frames.
  3. Send once, with a capture running.
  4. Read back.
  5. Write the result into the reference.

Order: stop/start → call-home off/on → setup push of one field → setup push of several fields.

Series-3 session test (before Series-3 sessions are built): hold a call-up session open to a MiniMate that is monitoring, for several minutes. Record:

  • does it keep monitoring (does an event still record)?
  • after hangup, is it monitoring?
  • does it behave differently on an RV55 with the DCD issue?

Build phases

Each safety piece arrives with its first real use, so nothing sits untested.

  1. One contract, reads only. device_access.py, the two adapters, the shared models, series selection, the per-unit lock, the error mapping, and sessions with ring safety. Rows 1–5 for both series through the same endpoints. No writes. Split into three plans:
    • 1a. The foundation plus identify / status / settings for both series.
    • 1b. Event list and download for both series (rows 4–5). Decided with Brian 2026-10-21:
      • MiniMate events are listed from their headers (0A + 1E arm + 0C + 1F browse, no 5A) instead of downloading every event to list it, and one event downloads without the ones before it. It reuses the bypass skip_waveform_for_events already takes, so no new command reaches the wire. The legacy serial-less /device/events keeps its full-download walk. 🔶 Unproven on a unit: a MiniMate bench check is owed before this branch merges. Its scope (final review, 2026-10-21):
        • a list of three or more events, confirming 1F(browse) advances after a 1E(arm) without 1F(arm);
        • a non-first event;
        • a histogram-mode event, capturing its 0C (which would let Series 3's record_type, null today, be decoded);
        • an event past 64 KB, expecting the named 502;
        • a stale-id download, confirming no 5A goes out;
        • a download of an event Blastware's call-home already delivered (one waveform, one histogram), expecting already_stored. This is the only way to measure that Series 3's (serial, timestamp) agrees between the unit's 0C and the watcher's import; if it does not, a download silently stores a duplicate row (never a downgrade). Measured for Series 4: 1,016/1,016 watcher imports found by the lookup (2026-10-21).
        • a dropped reply, via bridges/mm_link.py-style fault injection.
      • Download = fetch, check, store in SFM's database, the same as Series 3's blastware_file route, and returns the stored row's id. The file and waveform are then served by the existing /db/events/{id}/… routes. The database matches on (serial, timestamp), so storing an event already stored, including one THOR's watcher forwarded, refreshes the row instead of duplicating it. Changed by the final review (2026-10-21): it returns the existing row and writes nothing (inserted: false, detail.already_stored: true). Never merge: the watcher's copy, with its paired report, is the richer one, and a refresh measurably downgraded it (PVS, project strings and ZC frequencies went NULL).
      • event_id = <serial>:<key>:<YYYYmmddHHMMSS>. Both series reuse keys (after an erase, and across Micromates), so a download whose key now holds a different event fails with 409 instead of fetching it.
      • Micromate downloads walk to the event first (the only order captured from THOR), with THOR's 1,024-byte chunks.
    • 1c. The webapp: serial + address, and the same read panels for both series.
      • HARD PREREQUISITE: before any poller or webapp sends serial, add a per-address cool-down after a RADIO RING: call-ups to that address answer 503 + Retry-After until it expires, and POST /device/hangup clears it. The length must come from the unit's measured call-home retry timing (a bench test). Without it a poller can re-grab the modem's single TCP session through every retry, and the call never reaches THOR.
  2. Encoder + opcode guard + dry run + frame-reproduction tests. Still nothing sent. (2026-10-22) "Dry run" in this phase means the write plan: the exact frames, in order, with their expected acks, built from the unit's own reads (micromate.writes), plus a bench tool that prints one from a live unit's reads (bridges/mm_dry_run.py). Each write endpoint's dry_run=true arrives with that write in phases 3-5 and returns the plan. The opcode guard is a send-side gate in MicromateProtocol._send: every write SUB is refused unless the connection was built with it allowed (none is, in this phase), and 0xA2 is refused always.
  3. Start/stop monitoring (row 7): the first live write, at the bench. Brings the identity check, the audit table and the read-back framework.
  4. Call-home enable/disable (row 9).
  5. Setup push (row 10): one field first, then several.

Each phase is its own implementation plan. All phases land on feat/series4-sfm-parity and merge to dev once, when everything here is implemented (Brian, 2026-10-21) — not one merge per phase.


Out of scope (this phase)

  • Series-4 emergency tools: runaway stops, rescue, per-event delete, erase. (Erase-all is also unmapped.)
  • Async jobs; auth (Terra-View as jump host); the call-home receiver. These are shared features for later.
  • Scheduler read/write/enable; clock set; remote sensor check; restart; firmware.
  • Series-3 sessions, pending the bench test.
  • Mapping the rest of 0x2C (the THOR round).

Fix list (found 2026-10-19, separate small changes)

  • bridges/ach_server.py:524: seen_keys is undefined. Series-3 call-home monitor-log collection has failed every session since 9afa348 (2026-05-07).
  • bridges/mm_ach_server.py:262: _cstring/_content aren't imported, so the ring-answer path would raise NameError. From 8f67608 (2026-10-14).
  • Series-3 5A 64 KB page-boundary bug (known; surfaces as HTTP 500).
  • POST /device/config writes a fixed 29-byte 0x82 blob from one 3-11-26 capture, whatever the unit's trigger settings.
  • /device/events runs a full 5A download per event just to list headers.
  • 0x08 is named read_event_index in micromate/; it is the preferences block.

Left open by the phase 1a review (2026-10-21)

Minor findings from the whole-branch review, parked here so they outlive the plan's scratch ledger. None blocks 1a; the phase each one matters to is named.

  • CORS ordering (1c, if a browser calls SFM directly). The /device guard middleware is registered after CORS, so it sits outside it: its 409/501 responses carry no access-control-allow-origin, and an OPTIONS preflight to a guarded URL gets the guard's refusal instead of CORS's 200. Register CORS last, or let OPTIONS through. Terra-View proxies server-side and is not affected.
  • Legacy handlers take no unit lock (1c / phase 3). The guard checks for a held line and then lets the legacy handler run unlocked; a legacy request admitted just before a Series-4 call-up at the same address can overlap it.
  • Address keys are not normalised. tcp:<host>:<port> is the raw string: a hostname vs an IP, or /dev/serial/by-id/... vs /dev/ttyACM0, gets two locks for one link.
  • A serial-less request to an UNHELD Micromate still runs Series-3 bytes. The guard only knows about lines SFM holds. This is the phase-2 opcode guard's job.
  • tcp_port coercion mismatch. The guard parses with int(), pydantic coerces more loosely ("9034.0"), so such a request passes the guard unchecked. Parse with pydantic's own int coercion, or fail closed.
  • Test gaps. The middleware wrapper itself is untested (a raw ASGI call works without httpx); the contract test compares field names of one shared dataclass, which cannot fail, and does not enforce "a field absent for one series lives under detail".
  • _ensure_reaper race. Two first Series-4 requests at once can start two reaper threads; harmless because try_hold keeps them off each other's units.
  • Held serial lines share /dev/ttyACM0 silently. pyserial's open is not exclusive, so a bench script beside a held line interleaves bytes with the keep-alive; it looks like a dead unit. Open with exclusive=True, or don't hold serial links.

Left open by the phase 1b review (2026-10-21)

Parked from the whole-branch review of 1b; none blocks it. The phase each matters to is named.

  • Retry re-runs a whole download (1c). The retry-once on a held-line timeout re-runs the full download, roughly doubling time-to-error; narrow it to failures before the first reply.
  • last_used after a kept-line error (1c). It is not refreshed when an op's verdict (404/409) keeps the line, so the idle hang-up comes early.
  • verified: null is ambiguous (1c). It means both "a histogram" (no check exists) and "a waveform whose check could not run".
  • Mic naming and the Series-4 mic decode (1c). detail.mic_psi on Series 3, detail.mic on Series 4; Series 4's 0x0C mic decodes as a denormal (~5e-39) — fix the decode before 1c shows it.
  • 500 detail leaks paths (1c, before SFM is reachable beyond Terra-View). The download handler's 500 detail includes raw exception text, filesystem paths included.
  • Serial as a directory name (1c). The serial is never checked against [A-Z0-9]+ before it names a store directory; only the unit can choose it, but the check is cheap.
  • No server-level RADIO RING mid-download test (1c). Covered structurally (DeviceAccess-level test + the ring transport), not through the handler.
  • Series-3 stored record_type (MiniMate bench check). Rows SFM stores for Series 3 still say "Waveform" for histograms — pre-existing, the same as the legacy blastware_file route; fix it with the 0C decode.
  • A fresh Series-4 download's gaps (1c). No sample_rate is stored, and the sidecar has no vector sum; the DB row has the unit's PVS (overlaid from 0x0C).
  • Orphan files (SDM). An event file in the store with no DB row is still overwritten by a download.
  • Same-second collisions (SDM). Two events of one unit in the same second collide on (serial, timestamp) — a DB-wide dedup property, not 1b's.
  • 423 without Retry-After (1c). Add Retry-After to the 423 a request gets while a download holds the unit.
  • A lost 0C during a Series-3 list (1c). The strict walk raises on a lost 0A or 1F, but a lost 0C still lists the event with no timestamp and an event_id ending :-. Downloading that id skips the stale check and runs the full 5A. If the 0C is lost again it is refused (no timestamp, no filename); if it reads, whatever event now sits at that key is stored with no stale check: self-consistent data, but possibly not the event that was listed. Fix: raise on a lost 0C in strict mode, or refuse :- ids for Series 3.
  • A lost reply on a Series-3 TCP walk re-runs it once (1c) under the adapter's cold-boot retry before the 504, doubling time-to-error on a bad link.
  • A Series-3 download of an already-stored event still runs the full 5A before the database lookup finds it (the lookup is on the SDM side, after the device layer). Wasted transfer, not bad data (1c).

Table 2: to be added later (both series)

area item
infrastructure async jobs; auth via Terra-View
THOR has it, SFM doesn't clock set; remote sensor check; scheduler; one setup to many units; restart
call-home receiver hold lines open (S4); "already downloaded" tracking; rules by serial prefix (copy events, monitor log, delete, set clock); detect "Monitor While Calling Home" blocking deletes (⚠ one unexplained 21 → 5 count drop in THOR's log); hand off to /db/import/* instead of the watchers
settings, the no-PC-files way intended settings per deployment; drift report (intended vs actual); push then verify
notifications email/SMS by event type and PPV (THOR does it; may belong in Terra-View)
emergency tools for S4 if a Micromate ever needs them

Parking lot

Monitor logs. Its own piece of work, data side as much as device side. A unit calls in at least once or twice a day. With no events (e.g. waveform mode, no blast that day) it still delivers a monitor log: "monitoring from X to Y". The watchers (one for BW-ACH / Series 3, one for THOR-ACH / Series 4) use the newest file per unit, event or .MLG, as SFM's "last heard from" age. They forward waveform and histogram files but not monitor logs. To decide:

  • Ingest call-in monitor logs into SFM. SFM's DB already has a monitor_log table (one row per interval, deduped by serial + start time), and minimateplus/blastware_file.py already parses Series-3 .MLG. Missing: anything that takes .MLG in from the watchers, and knowledge of THOR's Series-4 monitor-log file format.
  • A live monitor-log read over a connection (Series 4: 0x0A exists, no wrapper or decoder; Series 3 in SFM: none).
  • What it is for: proof of coverage ("was it monitoring when they blasted?") and visible gaps.
  • Related fix-list item: ach_server.py's seen_keys.

How SFM handles setup files and schedules. This phase only reads and pushes the unit's active setup, with no PC-side library (the .mmb same-name/different-contents gotcha). Still to decide, as one question because the two are coupled:

  • The unit stores up to 25 named .MMB files. Should SFM manage them (list, select, rename, delete, push to a named file), or only ever touch the active one?
  • Schedules name setup files. A schedule entry switches the unit to a setup by name, so a schedule is only as reliable as the files it points to. What if it points to a file that isn't there? That is untested.
  • How "intended settings per deployment" (Table 2) fits with files and schedules.
  • Scheduler read/write is mostly mapped (file transfer + 0x47); disable has never been captured.

A modem-side agent on the RX55. AirLink OS on the RX55 has "Container Applications (Early Adopter)", and Brian already runs slmm-proxy there. A small program owning the unit's serial port could:

  • let THOR and SFM share the line
  • dial out to SFM, avoiding the IP whitelist
  • handle RADIO RING locally
  • Brian's hunch: make RX55s work with Series 3 by driving DTR/DCD itself

Open questions:

  • Can a container open the RS-232 port?
  • Can it set the control lines?
  • How is it deployed to a fleet, and does it Autostart?
  • Will a firmware update keep the beta feature?

Open questions this design leaves for the bench

  • The ack for start/stop when the unit is already in that state.
  • The offset rule, and whether 0x68/0x82 are required, for the setup push.
  • Whether the device ever refuses a write while monitoring. Only THOR's policy has been seen.
  • How a failed write is signalled. It has never been observed.
  • MiniMate behaviour with a held session (above).