# SFM: Series-4 parity and a device safety layer — design
|
||
|
||
**Date:** 2026-10-20 · **Branch:** `feat/series4-sfm-parity` · **Status:** draft for review
|
||
**Decided with:** Brian, in conversation 2026-10-19/20.
|
||
|
||
## In plain terms
|
||
|
||
SFM can already *read* a Micromate almost as well as it reads a MiniMate, but
|
||
only through the Python library and bench scripts. SFM's web endpoints still
|
||
only speak Series 3. This design does three things:
|
||
|
||
1. Gives Series 4 the same everyday jobs in SFM that Series 3 has: identify a
|
||
unit, check its status, read its settings, list and download events,
|
||
start/stop monitoring, and change its settings.
|
||
2. Adds a **safety layer** every device request goes through. It makes sure a
|
||
command only reaches the unit it was meant for, logs every change, and
|
||
checks every change by reading the unit back.
|
||
3. Lets SFM **stay on the line** with a Micromate. It calls up once, confirms
|
||
the serial, and keeps the connection open for later commands instead of
|
||
reconnecting each time.
|
||
|
||
**The guiding principle (Brian): talking to a Series 3 unit and talking to a
|
||
Series 4 unit is the same for the user.** It is the same screens, the same
|
||
endpoints, the same way of naming a unit, and the same shape of answer. The
|
||
only visible difference is the serial (`UM…` vs `BE…`) and the settings each
|
||
model has. The differences live underneath, in one adapter per series.
|
||
|
||
It does **not** cover the emergency tools (runaway-unit stops, rescue, erase),
|
||
the call-home receiver, or anything async. Those come later, as features
|
||
shared by both series.
|
||
|
||
**Production reach, stated honestly:** while THOR is set to stay connected to
|
||
production Micromates, a modem's single connection is taken. SFM can't reach
|
||
those units until THOR lets go or SFM takes over call-home. In this phase,
|
||
Series-4 SFM is for the bench and for units THOR is not holding.
|
||
|
||
---
|
||
|
||
## Decisions
|
||
|
||
| decision | why | who |
|
||
|---|---|---|
|
||
| **One user experience for both series** | the user's method is identical; only the serial prefix and each model's settings differ; adapters absorb the rest | Brian |
|
||
| Match **jobs**, not endpoints | some Series-3 endpoints only exist to work around Series-3 quirks | agreed |
|
||
| **No Series-4 emergency tools** yet (runaway stop, rescue, erase) | Micromates rarely need them | Brian |
|
||
| The series comes from the **serial prefix**: `UM` = Micromate, `BE` = MiniMate Plus, `BA` = Blastmate | the prefixes are definitive, so there is no probing | Brian |
|
||
| Before any **write**, confirm the unit that answered is the one named | SFM addresses units by IP; a moved modem or stale IP means a different unit answers | agreed |
|
||
| **Synchronous** endpoints for now; async later, for both series together | parity first, then shared features | agreed |
|
||
| **No library of setup files on the PC** (unlike THOR). The unit's own settings are the truth: SFM reads them from the unit, and verifies every push by reading the unit back | THOR keeps one copy of each setup on the PC and another on the unit, linked only by file name. Nothing checks the contents match, so the PC's `test2` and the unit's `test2` can say different things | Brian |
|
||
| **Sessions for Series 4 now; Series 3 only after a bench test** | THOR holds Micromate lines daily without harm; a MiniMate's hangup/DCD behaviour is untested | Brian |
|
||
| Audit log in **SFM's database** (new table) | Terra-View can show it; it fits the existing `CREATE TABLE IF NOT EXISTS` pattern | proposed |
|
||
|
||
---
|
||
|
||
## Scope: the parity table
|
||
|
||
Rows 8 and 11 of the original inventory (runaway stops, rescue/erase) are out
|
||
of scope by decision. Row 6 (monitor log) moved to the parking lot: it is its
|
||
own piece of work, together with call-in monitor-log ingest.
|
||
|
||
| # | job | Series 3 in SFM today | Series 4: protocol status | Series 4 in this phase |
|
||
|---|---|---|---|---|
|
||
| 1 | Identify a unit | `/device/info` (walks the whole chain, slow) | implemented: `connect()` | dispatch to the Micromate client |
|
||
| 2 | Quick status (monitoring, battery, memory, clock) | `monitor/status`, `storage_range`, `events/index` | implemented: `get_state()`; the `0x06` event count is decoded only in the bench tool | same endpoints; move the count decode into the client |
|
||
| 3 | Read settings | compliance in `/device/info`; `GET /device/call_home` | `get_setup()` implemented and verified on hardware; call-home: enable + dial string observed, rest unmapped | setup in full; call-home returns enable + dial string + raw |
|
||
| 4 | List events | `/device/events` (downloads every event to list them) | implemented, and cheap: the chain walk returns sizes | the same event-list shape as Series 3, each event carrying an `event_id` |
|
||
| 5 | Download an event | `/event/{i}`, `/waveform`, `/blastware_file` | implemented: streams the `.IDFW`/`.IDFH` itself, with a decode self-check | the same download and save-to-DB actions as Series 3. Underneath, Series 4 saves through `/db/import/idf_file` |
|
||
| 7 | Start / stop monitoring | `monitor/start`, `monitor/stop` | mapped, identical to Series 3 (`0x96`/`0x97`, acks `0x69`/`0x68`); never sent | **first write.** Verify with `0x49` state |
|
||
| 9 | Change call-home settings | `POST /device/call_home` | write mapped (`0x2C` → `0x7E`/`0x7F`, same as Series 3); field map mostly unknown | enable/disable only; the rest waits for the THOR round |
|
||
| 10 | Change recording settings | `POST /device/config` | sequence mapped (`0xDA` → `0x68/73` → `0x82/83` → `0x71/72`); no encoder | read-modify-write, stop → push → restart, verify with `get_setup()` |
|
||
|
||
---
|
||
|
||
## Architecture
|
||
|
||
### One door for every device request
|
||
|
||
A new module, `sfm/device_access.py`, sits between the `/device/*` handlers and
|
||
the two device libraries. No handler opens a transport or picks a library
|
||
itself any more.
|
||
|
||
```
|
||
/device/* handler
|
||
│ (serial, address, action, tier)
|
||
▼
|
||
device_access ── series from serial prefix ── session / per-unit lock
|
||
│ │ │
|
||
│ ▼ ▼
|
||
│ minimateplus (S3) or micromate (S4) audit table
|
||
▼
|
||
read-back verification (changes only)
|
||
```
|
||
|
||
- **Naming a unit is the same for both: `serial` + address.** The prefix picks
|
||
the adapter: `UM` → Series 4 (`micromate`); `BE` or `BA` → Series 3
|
||
(`minimateplus`); any other prefix → 422. A request with no `serial` is
|
||
treated as Series 3. That is a **legacy path** kept only so existing scripts
|
||
keep working; the webapp and every new caller always send `serial`.
|
||
- **One contract.** Same `/device/*` paths, same request parameters, and
|
||
**the same response models for both series**:
|
||
- `UnitInfo`, `UnitStatus`, `UnitSettings`, `EventRef`, `ChangeResult`
|
||
- Anything only one series has goes under a `detail` object inside the same
|
||
model, never as a different top-level shape. `"series": 3 | 4` is
|
||
informational.
|
||
- **Events have one identity.** An opaque `event_id` string from the event
|
||
list, passed back to download it. Underneath it is Series 3's position or
|
||
Series 4's device key, plus the serial; the user never handles either.
|
||
- **Not-yet-supported is uniform.** If one series can't do something yet,
|
||
the endpoint still exists and answers **501 "not supported for Series N
|
||
yet"**. It is never a missing route or a different call.
|
||
- **Work separate from HTTP.** Each operation is a plain function taking the
|
||
session. The handler is a thin wrapper, so making an operation an async job
|
||
later means changing the wrapper, not the operation.
|
||
|
||
### Adapters
|
||
|
||
Each series implements the same operation set behind `device_access`:
|
||
|
||
| operation | Series 3 adapter | Series 4 adapter |
|
||
|---|---|---|
|
||
| `identify` | STARTUP + serial + config | `connect()` |
|
||
| `status` | `0x1C` (+ `0x06`) | `get_state()` + `0x06` count |
|
||
| `read_settings` | `0x1A` compliance decode | `get_setup()` |
|
||
| `list_events` | today's walk | chain walk (cheap) |
|
||
| `download_event` | 5A → Blastware file | 5A → `.IDFW`/`.IDFH` |
|
||
| `start` / `stop` | `0x96` / `0x97` (exists) | `0x96` / `0x97` (new) |
|
||
| `read_call_home` / `set_call_home_enabled` | exists | `0x2C` / `0x7E`+`0x7F` |
|
||
| `push_settings` | `/device/config` path (exists) | setup push (new) |
|
||
| `call_up` / `hang_up` | connect-per-request underneath, until the bench test | held session |
|
||
|
||
Each adapter maps its native data into the shared models. `UnitSettings` keeps
|
||
the fields both series have under the same names and units: record mode,
|
||
sample rate, record time, histogram interval, trigger levels, the notes as
|
||
label → value. Model-specific fields go under `detail`.
|
||
|
||
**Requests that include `serial` get the shared models; requests without it
|
||
keep today's Series-3 shapes, unchanged.** Nothing breaks in between: the
|
||
webapp moves to `serial` in phase 1c, and the old path retires after that and
|
||
after `watch_unit.sh` moves. Terra-View does not call `/device/*`. The
|
||
emergency endpoints (`stop_monitoring_*`, `rescue`, `events/erase`) are not
|
||
touched and stay Series-3 only; a `UM` serial gets 501.
|
||
|
||
### The webapp
|
||
|
||
The same screens for both series. You name a unit by serial + address (today
|
||
it takes host/port only), and the series shows as a badge, nothing more. Read
|
||
panels for both series land in phase 1; each write button appears for Series 4
|
||
in the phase that adds that write.
|
||
|
||
### The safety layer
|
||
|
||
**Tiers.** Every operation declares one:
|
||
|
||
| tier | examples | what the layer does |
|
||
|---|---|---|
|
||
| read | info, status, setup, events, download | runs; no identity check |
|
||
| change | start/stop, call-home, setup push | identity check → audit → send → **read back** → audit the result |
|
||
| destructive | delete, erase (none in this phase) | everything above + the caller must send the unit's serial as `confirm` |
|
||
|
||
**Identity check (changes only).** Before the first change in a connection,
|
||
compare the serial the unit reports with the one requested; on a mismatch,
|
||
refuse (409) and send nothing. Series 4 already reads the serial in
|
||
`connect()`. A Series-3 change adds one serial read (`SUB 0x15`), about one
|
||
round trip.
|
||
|
||
**Opcode guard.** A send-side check that refuses known cross-series
|
||
collisions. It starts with **`0xA2`**, which is CONFIRM ERASE on Series 3 and
|
||
"end call-home" on Series 4. This phase never sends `0xA2`; the guard makes
|
||
that impossible by accident. The table of colliding opcodes lives in one
|
||
place.
|
||
|
||
**One connection per unit.** A per-unit lock keyed by address
|
||
(`tcp:host:port` / `serial:port`). A request for a busy unit waits up to
|
||
**20 s** (configurable; under Terra-View's 60 s proxy timeout), then gets
|
||
**423 "unit busy"** instead of opening a second connection. This applies to Series 3 as well. The webapp's Diagnostics
|
||
refresh currently fires three device requests at once against a modem that
|
||
allows one; under the lock they queue.
|
||
|
||
**Read-back verification.** A zero-data ack has never been proof that a write
|
||
applied. Every change is followed by a read:
|
||
- start/stop → `0x49` state: start expects `0x02`, stop expects `0x00`. Record
|
||
the value either way; `0x03` is the in-call-home state
|
||
- call-home → re-read `0x2C` and compare the edited bytes
|
||
- setup push → `get_setup()` and compare every field that was set
|
||
|
||
A mismatch is reported as a failed change (and audited as one), even if the
|
||
ack was clean.
|
||
|
||
**Errors that say what happened.** Distinct responses for:
|
||
- 503: couldn't connect
|
||
- 504: connected but the unit didn't answer
|
||
- 423: unit busy
|
||
- 409: wrong unit
|
||
- 422: bad input
|
||
- 502: the reply was malformed
|
||
|
||
Today 502 covers both of the first two.
|
||
|
||
**Audit table.** New table `device_commands`, created on startup like the
|
||
others. It is additive, but still a schema change, so the changelog says so.
|
||
One row per change attempt:
|
||
|
||
| column | content |
|
||
|---|---|
|
||
| `ts`, `serial`, `series`, `address` | who and where |
|
||
| `action`, `tier` | e.g. `monitor_stop`, `change` |
|
||
| `request` | parameters, plus hex of what was sent |
|
||
| `ack` | reply summary |
|
||
| `verified` | read-back result: `ok` / `mismatch` / `not_checked` |
|
||
| `outcome`, `error` | final result |
|
||
| `caller` | best-effort origin (`webapp`, `terra-view`, `script`) |
|
||
|
||
Reads are not audited.
|
||
|
||
### Sessions
|
||
|
||
1. **Call up.** Connect, `connect()` (POLL / SERIAL / STATE), identity check
|
||
against the requested serial.
|
||
2. **On the line.** Requests for that unit run in turn on the same connection,
|
||
behind the per-unit lock.
|
||
3. **Keep-alive.** A POLL every 60 s while idle. The ACEmanager TCP idle
|
||
timeout is 2 min.
|
||
4. **Hang up.** On `POST /device/hangup`, after 5 idle minutes, or on any
|
||
transport error. A reconnect is a new session with a new identity check.
|
||
5. **Ring safety.** If `RADIO RING` arrives on an SFM session, log it and hang
|
||
up **immediately**. Until SFM is the call-home destination, the unit's
|
||
retry should reach THOR. 🔶 *Inferred:* the unit retries; its `0x2C`
|
||
retries field reads 3.
|
||
|
||
**Call up and hang up are the same for both series.** Underneath, the
|
||
Series-3 adapter keeps today's connect-per-request behaviour until the bench
|
||
test under *Testing*; the user sees the same actions either way.
|
||
|
||
### Series-4 write path
|
||
|
||
**Encoder.** Extend `micromate/framing.py`:
|
||
- a request builder that carries a data section
|
||
- the same escape set `{02,03,04,10}`, applied to data, params and checksum
|
||
- plain SUM8 of the destuffed payload (*not* Series 3's write format)
|
||
|
||
Rules: `micromate_protocol_reference.md` → *The escape set — exactly four bytes*,
|
||
*Checksum — plain SUM8 of the de-stuffed payload*, *Write-frame destuffing*.
|
||
|
||
**Dry run.** Every Series-4 write accepts `dry_run=true`. It returns the exact
|
||
frames it would send and sends nothing. The first live send of each write is
|
||
reviewed this way first.
|
||
|
||
**Per operation:**
|
||
|
||
| operation | frames | verify | open questions it settles |
|
||
|---|---|---|---|
|
||
| start / stop | `0x96` / `0x97`, offset `0xFFFF`, zero params | `0x49` | the ack when already in that state |
|
||
| call-home enable/disable | read `0x2C` → `0x7E` (126 B content, offset `0x007E`) → `0x7F` | re-read `0x2C` | none new; bit 0 of `write[5]` |
|
||
| setup push | read `0x08`, `0x2E`, `0x1A` → `0xDA` (name) → `0x68`/`0x73` → `0x82`/`0x83` → `0x71`/`0x72` | `get_setup()` | the offset rule; whether `0x68`/`0x82` are needed |
|
||
|
||
**Setup push, details:**
|
||
- Read-modify-write of **all three** blocks, as THOR appears to do.
|
||
🔶 *Inferred from two captures:* `0x68` = the preferences content with
|
||
`[0:3]` zeroed; `0x82` = the `0x2E` content with `[0:3]` and the CRC slot
|
||
zeroed; `0x71` = the `0x1A` content `[0:2090]` with the length echo zeroed.
|
||
- Offsets come from THOR's observed constants. 🔶 The hypothesis "offset =
|
||
the read block's length + 2" fits all four, but each has been seen at one
|
||
size only.
|
||
- The edited fields come from `MicromateSetup` and are written back at the
|
||
offsets `setup_block.py` already uses. Only fields graded **measured** may
|
||
be edited.
|
||
- Sequence: stop monitoring (if running) → push → restart (if it was running)
|
||
→ verify. THOR's "must be idle" is policy, not the device, but stopping
|
||
first removes a variable from the first live write.
|
||
- ⚠ **Edit-buffer hazard.** `0x1A` reads the keypad's unsaved edits, so a
|
||
read-modify-write taken while someone is in the menu would commit them.
|
||
This can't be detected from the wire; it is documented in the endpoint and
|
||
audited.
|
||
- **Target file:** the active setup, overwritten in place, by default. A
|
||
named target is allowed, but overwriting a *non-active* setup or
|
||
`factory.MMB` is untested, so it is refused in this phase.
|
||
|
||
---
|
||
|
||
## Testing
|
||
|
||
**Offline (CI-capable):**
|
||
- **Frame reproduction.** The new encoder must reproduce THOR's captured write
|
||
frames byte for byte: `0x96`/`0x97`, `0x7E`/`0x7F`, and the full setup push,
|
||
from the 9-24 / 9-25 / 10-14 captures. This is the same method that proved
|
||
the reads (218/218). The needed frames are copied into force-added fixtures,
|
||
since `bridges/captures/` is gitignored.
|
||
- **Safety layer on fakes:**
|
||
- series by prefix
|
||
- identity mismatch → 409, nothing sent
|
||
- `0xA2` refused
|
||
- lock → 423
|
||
- read-back mismatch → failed change
|
||
- audit row written for every change attempt
|
||
- RADIO RING → session ends
|
||
- **Endpoints** through FastAPI's `TestClient` against the scripted transport
|
||
already used in `tests/test_micromate_client.py`.
|
||
- **Contract tests.** The same test, run against both adapters with fakes,
|
||
asserts that each job returns the same model with the same field names and
|
||
units. A field present for one series and absent for the other is a test
|
||
failure unless it lives under `detail`. This is the test that keeps the
|
||
guiding principle true as the code grows.
|
||
|
||
**Bench (UM20147, with Brian):** each write's first live send follows the same
|
||
protocol:
|
||
1. Dry run.
|
||
2. Review the frames.
|
||
3. Send once, with a capture running.
|
||
4. Read back.
|
||
5. Write the result into the reference.
|
||
|
||
Order: stop/start → call-home off/on → setup push of one field → setup push of
|
||
several fields.
|
||
|
||
**Series-3 session test (before Series-3 sessions are built):** hold a call-up
|
||
session open to a MiniMate that is monitoring, for several minutes. Record:
|
||
- does it keep monitoring (does an event still record)?
|
||
- after hangup, is it monitoring?
|
||
- does it behave differently on an RV55 with the DCD issue?
|
||
|
||
---
|
||
|
||
## Build phases
|
||
|
||
Each safety piece arrives with its first real use, so nothing sits untested.
|
||
|
||
1. **One contract, reads only.** `device_access.py`, the two adapters, the
|
||
shared models, series selection, the per-unit lock, the error mapping, and
|
||
sessions with ring safety. Rows 1–5 for **both** series through the same
|
||
endpoints. **No writes.** Split into three plans:
|
||
- 1a. The foundation plus identify / status / settings for both series.
|
||
- 1b. Event list and download for both series (rows 4–5). Decided with
|
||
Brian 2026-10-21:
|
||
- **MiniMate events are listed from their headers** (`0A` + `1E` arm +
|
||
`0C` + `1F` browse, no `5A`) instead of downloading every event to list
|
||
it, and one event downloads without the ones before it. It reuses the
|
||
bypass `skip_waveform_for_events` already takes, so no new command
|
||
reaches the wire. The legacy serial-less `/device/events` keeps its
|
||
full-download walk. 🔶 Unproven on a unit: **a MiniMate bench check is
|
||
owed before this branch merges.** Its scope (final review,
|
||
2026-10-21):
|
||
- a list of three or more events, confirming `1F(browse)` advances
|
||
after a `1E(arm)` without `1F(arm)`;
|
||
- a non-first event;
|
||
- a histogram-mode event, capturing its `0C` (which would let Series
|
||
3's `record_type`, `null` today, be decoded);
|
||
- an event past 64 KB, expecting the named 502;
|
||
- a stale-id download, confirming no `5A` goes out;
|
||
- a download of an event Blastware's call-home already delivered (one
|
||
waveform, one histogram), expecting `already_stored`. This is the
|
||
only way to measure that Series 3's (serial, timestamp) agrees
|
||
between the unit's `0C` and the watcher's import; if it does not,
|
||
a download silently stores a duplicate row (never a downgrade).
|
||
Measured for Series 4: 1,016/1,016 watcher imports found by the
|
||
lookup (2026-10-21).
|
||
- a dropped reply, via `bridges/mm_link.py`-style fault injection.
|
||
- **Download = fetch, check, store** in SFM's database, the same as
|
||
Series 3's `blastware_file` route, and returns the stored row's id. The
|
||
file and waveform are then served by the existing `/db/events/{id}/…`
|
||
routes. The database matches on (serial, timestamp), so storing an
|
||
event already stored, including one THOR's watcher forwarded, ~~refreshes
|
||
the row instead of duplicating it~~. **Changed by the final review
|
||
(2026-10-21):** it returns the existing row and writes nothing
|
||
(`inserted: false`, `detail.already_stored: true`). Never merge: the
|
||
watcher's copy, with its paired report, is the richer one, and a
|
||
refresh measurably downgraded it (PVS, project strings and ZC
|
||
frequencies went NULL).
|
||
- **`event_id` = `<serial>:<key>:<YYYYmmddHHMMSS>`.** Both series reuse
|
||
keys (after an erase, and across Micromates), so a download whose key
|
||
now holds a different event fails with 409 instead of fetching it.
|
||
- **Micromate downloads walk to the event first** (the only order captured
|
||
from THOR), with THOR's 1,024-byte chunks.
|
||
- 1c. **The webapp:** serial + address, and the same read panels for both
|
||
series.
|
||
- **HARD PREREQUISITE:** before any poller or webapp sends `serial`, add a
|
||
per-address cool-down after a RADIO RING: call-ups to that address answer
|
||
503 + `Retry-After` until it expires, and `POST /device/hangup` clears it.
|
||
The length must come from the unit's measured call-home retry timing (a
|
||
bench test). Without it a poller can re-grab the modem's single TCP
|
||
session through every retry, and the call never reaches THOR.
|
||
2. **Encoder + opcode guard + dry run + frame-reproduction tests.** Still
|
||
nothing sent.
|
||
(2026-10-22) "Dry run" in this phase means the **write plan**: the exact
|
||
frames, in order, with their expected acks, built from the unit's own reads
|
||
(`micromate.writes`), plus a bench tool that prints one from a live unit's
|
||
reads (`bridges/mm_dry_run.py`). Each write endpoint's `dry_run=true`
|
||
arrives with that write in phases 3-5 and returns the plan. The opcode
|
||
guard is a **send-side gate** in `MicromateProtocol._send`: every write SUB
|
||
is refused unless the connection was built with it allowed (none is, in
|
||
this phase), and `0xA2` is refused always.
|
||
3. **Start/stop monitoring** (row 7): the first live write, at the bench.
|
||
Brings the identity check, the audit table and the read-back framework.
|
||
4. **Call-home enable/disable** (row 9).
|
||
5. **Setup push** (row 10): one field first, then several.
|
||
|
||
Each phase is its own implementation plan. **All phases land on
|
||
`feat/series4-sfm-parity` and merge to `dev` once, when everything here is
|
||
implemented** (Brian, 2026-10-21) — not one merge per phase.
|
||
|
||
---
|
||
|
||
## Out of scope (this phase)
|
||
|
||
- Series-4 emergency tools: runaway stops, rescue, per-event delete, erase.
|
||
(Erase-all is also unmapped.)
|
||
- Async jobs; auth (Terra-View as jump host); the call-home receiver. These
|
||
are shared features for later.
|
||
- Scheduler read/write/enable; clock set; remote sensor check; restart;
|
||
firmware.
|
||
- Series-3 sessions, pending the bench test.
|
||
- Mapping the rest of `0x2C` (the THOR round).
|
||
|
||
## Fix list (found 2026-10-19, separate small changes)
|
||
|
||
- `bridges/ach_server.py:524`: `seen_keys` is undefined. Series-3 call-home
|
||
monitor-log collection has failed every session since `9afa348`
|
||
(2026-05-07).
|
||
- `bridges/mm_ach_server.py:262`: `_cstring`/`_content` aren't imported, so
|
||
the ring-answer path would raise `NameError`. From `8f67608` (2026-10-14).
|
||
- Series-3 `5A` 64 KB page-boundary bug (known; surfaces as HTTP 500).
|
||
- `POST /device/config` writes a fixed 29-byte `0x82` blob from one 3-11-26
|
||
capture, whatever the unit's trigger settings.
|
||
- `/device/events` runs a full 5A download per event just to list headers.
|
||
- `0x08` is named `read_event_index` in `micromate/`; it is the preferences
|
||
block.
|
||
|
||
## Left open by the phase 1a review (2026-10-21)
|
||
|
||
Minor findings from the whole-branch review, parked here so they outlive the
|
||
plan's scratch ledger. None blocks 1a; the phase each one matters to is named.
|
||
|
||
- **CORS ordering (1c, if a browser calls SFM directly).** The `/device` guard
|
||
middleware is registered after CORS, so it sits outside it: its 409/501
|
||
responses carry no `access-control-allow-origin`, and an OPTIONS preflight to
|
||
a guarded URL gets the guard's refusal instead of CORS's 200. Register CORS
|
||
last, or let OPTIONS through. Terra-View proxies server-side and is not
|
||
affected.
|
||
- **Legacy handlers take no unit lock (1c / phase 3).** The guard checks for a
|
||
held line and then lets the legacy handler run unlocked; a legacy request
|
||
admitted just before a Series-4 call-up at the same address can overlap it.
|
||
- **Address keys are not normalised.** `tcp:<host>:<port>` is the raw string:
|
||
a hostname vs an IP, or `/dev/serial/by-id/...` vs `/dev/ttyACM0`, gets two
|
||
locks for one link.
|
||
- **A serial-less request to an UNHELD Micromate still runs Series-3 bytes.**
|
||
The guard only knows about lines SFM holds. This is the phase-2 opcode
|
||
guard's job.
|
||
- **`tcp_port` coercion mismatch.** The guard parses with `int()`, pydantic
|
||
coerces more loosely (`"9034.0"`), so such a request passes the guard
|
||
unchecked. Parse with pydantic's own int coercion, or fail closed.
|
||
- **Test gaps.** The middleware wrapper itself is untested (a raw ASGI call
|
||
works without httpx); the contract test compares field names of one shared
|
||
dataclass, which cannot fail, and does not enforce "a field absent for one
|
||
series lives under `detail`".
|
||
- **`_ensure_reaper` race.** Two first Series-4 requests at once can start two
|
||
reaper threads; harmless because `try_hold` keeps them off each other's units.
|
||
- **Held serial lines share `/dev/ttyACM0` silently.** pyserial's open is not
|
||
exclusive, so a bench script beside a held line interleaves bytes with the
|
||
keep-alive; it looks like a dead unit. Open with `exclusive=True`, or don't
|
||
hold serial links.
|
||
|
||
## Left open by the phase 1b review (2026-10-21)
|
||
|
||
Parked from the whole-branch review of 1b; none blocks it. The phase each
|
||
matters to is named.
|
||
|
||
- **Retry re-runs a whole download (1c).** The retry-once on a held-line timeout re-runs the full download, roughly doubling time-to-error; narrow it to failures before the first reply.
|
||
- **`last_used` after a kept-line error (1c).** It is not refreshed when an op's verdict (404/409) keeps the line, so the idle hang-up comes early.
|
||
- **`verified: null` is ambiguous (1c).** It means both "a histogram" (no check exists) and "a waveform whose check could not run".
|
||
- **Mic naming and the Series-4 mic decode (1c).** `detail.mic_psi` on Series 3, `detail.mic` on Series 4; Series 4's `0x0C` mic decodes as a denormal (~5e-39) — fix the decode before 1c shows it.
|
||
- **500 detail leaks paths (1c, before SFM is reachable beyond Terra-View).** The download handler's 500 detail includes raw exception text, filesystem paths included.
|
||
- **Serial as a directory name (1c).** The serial is never checked against `[A-Z0-9]+` before it names a store directory; only the unit can choose it, but the check is cheap.
|
||
- **No server-level RADIO RING mid-download test (1c).** Covered structurally (DeviceAccess-level test + the ring transport), not through the handler.
|
||
- **Series-3 stored `record_type` (MiniMate bench check).** Rows SFM stores for Series 3 still say "Waveform" for histograms — pre-existing, the same as the legacy `blastware_file` route; fix it with the `0C` decode.
|
||
- **A fresh Series-4 download's gaps (1c).** No `sample_rate` is stored, and the sidecar has no vector sum; the DB row has the unit's PVS (overlaid from `0x0C`).
|
||
- **Orphan files (SDM).** An event file in the store with no DB row is still overwritten by a download.
|
||
- **Same-second collisions (SDM).** Two events of one unit in the same second collide on (serial, timestamp) — a DB-wide dedup property, not 1b's.
|
||
- **423 without `Retry-After` (1c).** Add `Retry-After` to the 423 a request gets while a download holds the unit.
|
||
- **A lost `0C` during a Series-3 list (1c).** The strict walk raises on a lost `0A`
|
||
or `1F`, but a lost `0C` still lists the event with no timestamp and an
|
||
`event_id` ending `:-`. Downloading that id skips the stale check and runs the
|
||
full `5A`. If the `0C` is lost again it is refused (no timestamp, no filename);
|
||
if it reads, whatever event now sits at that key is stored with no stale check:
|
||
self-consistent data, but possibly not the event that was listed. Fix: raise on a lost `0C` in strict mode, or
|
||
refuse `:-` ids for Series 3.
|
||
- **A lost reply on a Series-3 TCP walk re-runs it once (1c)** under the adapter's
|
||
cold-boot retry before the 504, doubling time-to-error on a bad link.
|
||
- **A Series-3 download of an already-stored event still runs the full `5A`**
|
||
before the database lookup finds it (the lookup is on the SDM side, after the
|
||
device layer). Wasted transfer, not bad data (1c).
|
||
|
||
## Table 2: to be added later (both series)
|
||
|
||
| area | item |
|
||
|---|---|
|
||
| infrastructure | async jobs; auth via Terra-View |
|
||
| THOR has it, SFM doesn't | clock set; remote sensor check; scheduler; one setup to many units; restart |
|
||
| call-home receiver | hold lines open (S4); "already downloaded" tracking; rules by serial prefix (copy events, monitor log, delete, set clock); detect "Monitor While Calling Home" blocking deletes (⚠ one unexplained 21 → 5 count drop in THOR's log); hand off to `/db/import/*` instead of the watchers |
|
||
| settings, the no-PC-files way | intended settings per deployment; drift report (intended vs actual); push then verify |
|
||
| notifications | email/SMS by event type and PPV (THOR does it; may belong in Terra-View) |
|
||
| emergency tools for S4 | if a Micromate ever needs them |
|
||
|
||
## Parking lot
|
||
|
||
**Monitor logs.** Its own piece of work, data side as much as device side.
|
||
A unit calls in at least once or twice a day. With no events (e.g. waveform
|
||
mode, no blast that day) it still delivers a **monitor log**: "monitoring from
|
||
X to Y". The watchers (one for BW-ACH / Series 3, one for THOR-ACH / Series 4)
|
||
use the newest file per unit, event or `.MLG`, as SFM's "last heard from" age.
|
||
They forward waveform and histogram files but **not** monitor logs. To decide:
|
||
- Ingest call-in monitor logs into SFM. SFM's DB **already has** a
|
||
`monitor_log` table (one row per interval, deduped by serial + start time),
|
||
and `minimateplus/blastware_file.py` already parses Series-3 `.MLG`.
|
||
Missing: anything that takes `.MLG` in from the watchers, and knowledge of
|
||
THOR's Series-4 monitor-log file format.
|
||
- A live monitor-log read over a connection (Series 4: `0x0A` exists, no
|
||
wrapper or decoder; Series 3 in SFM: none).
|
||
- What it is for: proof of coverage ("was it monitoring when they blasted?")
|
||
and visible gaps.
|
||
- Related fix-list item: `ach_server.py`'s `seen_keys`.
|
||
|
||
**How SFM handles setup files and schedules.** This phase only reads and
|
||
pushes the unit's *active* setup, with no PC-side library (the `.mmb`
|
||
same-name/different-contents gotcha). Still to decide, as one question because
|
||
the two are coupled:
|
||
- The unit stores up to 25 named `.MMB` files. Should SFM manage them (list,
|
||
select, rename, delete, push to a named file), or only ever touch the active
|
||
one?
|
||
- **Schedules name setup files.** A schedule entry switches the unit to a
|
||
setup by name, so a schedule is only as reliable as the files it points to.
|
||
What if it points to a file that isn't there? That is untested.
|
||
- How "intended settings per deployment" (Table 2) fits with files and
|
||
schedules.
|
||
- Scheduler read/write is mostly mapped (file transfer + `0x47`); **disable has
|
||
never been captured**.
|
||
|
||
**A modem-side agent on the RX55.** AirLink
|
||
OS on the RX55 has "Container Applications (Early Adopter)", and Brian already
|
||
runs `slmm-proxy` there. A small program owning the unit's serial port could:
|
||
- let THOR and SFM share the line
|
||
- dial out to SFM, avoiding the IP whitelist
|
||
- handle `RADIO RING` locally
|
||
- **Brian's hunch:** make RX55s work with Series 3 by driving DTR/DCD itself
|
||
|
||
Open questions:
|
||
- Can a container open the RS-232 port?
|
||
- Can it set the control lines?
|
||
- How is it deployed to a fleet, and does it Autostart?
|
||
- Will a firmware update keep the beta feature?
|
||
|
||
## Open questions this design leaves for the bench
|
||
|
||
- The ack for start/stop when the unit is already in that state.
|
||
- The offset rule, and whether `0x68`/`0x82` are required, for the setup push.
|
||
- Whether the device ever refuses a write while monitoring. Only THOR's policy
|
||
has been seen.
|
||
- How a failed write is signalled. It has never been observed.
|
||
- MiniMate behaviour with a held session (above).
|