Brian's intent (UM6047's trigger, notes typos, the backup unit), the
flow, the first-version fields, and what is measured: Series 3's
compliance block has Series 4's 0x1A layout; a whole 1A read assembles
to the 2,090 bytes Blastware writes back (exact on both 10-26 writes);
Blastware writes 00 00 at [26:28] and a per-unit constant after the
block.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Blastware prints OORANGE for an interval's vector sum when it exceeds
the range in an interval where Blastware also clamped a channel: 17 / 17
DL2 cells, SFM's sum 10.01 .. 17.33 in/s (sqrt 3 x 10.005). The
saturation carve-out now covers intervals.Geo PVS in/s with that rule,
per interval, as it already did for PeakVectorSum. The 4 other cells
left mismatching sit 4e-6 past a rounding boundary (Blastware's
arithmetic is a hair lower) and stay mismatches.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Group B of the file-only decode, measured on DL2 2026-11-03:
- Frequency = (sample rate / 2) / half-period: 527,915 / 527,915 cells
at 1024 sps and 1,725 / 1,725 at 2048. #50: half_period_to_hz took a
fixed 512 and halved every 2048-sps frequency; it now takes the rate
(default 1024). Blastware prints 'N/A' for a half-period of 0, '>100'
('>200' at 2048) at 5 samples or fewer, '<1.0' below 1 Hz.
- Each interval's vector sum, sqrt(PVS^2) at the event's range:
771,354 / 771,354.
- Interval size: uint16 seconds at k - 40 (Series IV [8:10]).
- Stop: the footer's (2,995 / 2,999 sampled; 4 print a second earlier).
Event gains histogram_interval_sec, interval_half_periods, interval_pvs
and stop_time; the .h5 gains /intervals/half_period/<ch>,
/intervals/pvs_ips and a histogram_interval_sec attr. The decoder walks
a histogram once (histogram_codec.channels_from_records) and takes the
event's PVS from the same records. The checker compares each interval
column, number or Blastware's word; the test that pinned Tran Freq Hz
as blank now pins the match.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Protocol reference:
- D.6.8 (new): the identity block (SUB FE's: serial, firmware, DSP version,
calibration date and 'by', model; battery), the trigger-source bitmask
in the embedded event record, and the trigger settings (sources, stop
mode, per-channel levels, mic type) with their DL2 counts. The
histogram 'terminal tail' 9c 06 00 42 looks like 9c 06 00 + the
serial's B (inferred).
- D.6.6: recorded data vs report units vs unit peaks, with the
MiniMate Plus (s3.9.7) and Blastware (s4) manual sections. Where the
unit stores its own Microphone Units is unmeasured; a keypad test is
proposed.
- D.6.9 (was D.6.7): refreshed, 28 of 88.
CLAUDE.md 'Where things stand' and the README roadmap follow.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Both settings are documented: the unit's Microphone Units (MiniMate Plus
manual s3.9.7, keypad Right Arrow + Setups: psi(L) / dB(L) / Pa(L)) and
Blastware's report preference (Blastware manual s4, Preferences > Units
and Channel). The report follows Blastware's, as DL2's fleet-wide
switches showed. Where the unit's own setting is stored is unmeasured.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Brings be5bab4 and its history from the decode checker's branch: a Blastware
file's job number; decode_blastware_bytes choosing the codec by STRT[18] and the
histogram layout check by peak <= PVS <= sqrt(3) peak; the footer as the file's
last 26 bytes (no day-late histograms); docs (reference Appendix D.6, the
file-only status).
Conflicts, both kept whole:
- CLAUDE.md: this branch's 'Narrowed 2026-10-26' 5A block stays, followed by
dev's 'Open, not blocking' line (the 8x sensitive-range item struck out --
resolved by the file's range).
- docs/instantel_protocol_reference.md (CRLF, byte-safe): the change-log rows in
date order -- dev's 2026-11-03 and 2026-11-02, this branch's 2026-11-03
(call home), 2026-10-29 and 2026-10-27 (#46, this branch only).
Suite on the merged tree: 1399 passed, the 16 known failures.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Benched 2026-11-03 on UM20147 and BE9441 (each on its own modem, one send per
go): idle, a no-op, off and on all acked, read back and confirmed on each unit's
menu; UM20147 accepts 0x00 at [5]. Monitoring, neither series takes a change: a
MiniMate answers 0x2C with its device-info block, a Micromate acks and ignores
('Feature disabled while monitoring'). Series 4 [118:120] is a function of the
block's content -- XOR 0x6D7D per bit-0 flip on two units, not a nonce. A stop's
call-in reached THOR when the ring followed the ack. UM20147's clock is an hour
ahead (DST). Spec, both protocol references (Series III: a CRLF row), CLAUDE.md.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- The page's top bar Firmware read '—' for a Micromate: Batch 2 gave the Device
tab's card firmwareLabel() but not populateDeviceBar. Seen at the bench
2026-11-03; checked in jsdom ('—' before, 'Thor line (version not read)' after).
- save_imported_bw's overlay comment said the waveform codec was undecoded and
gave +-10 in/s; the overlay now stays because production keeps a paired
Blastware report as the record until the file-only decode has no blanks
(flagged by the decode-checker session). Comment only.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Measured at the bench 2026-11-03 (UM20147): monitoring, a Micromate returns its
real call-home block and ACKS a call-home write, but does not apply it -- three
read-backs unchanged, and its own menu says 'Feature disabled while monitoring'.
SFM now refuses up front (UnitStateUnknown, 409: '... is monitoring -- a Micromate
ignores call-home changes while monitoring ...; stop monitoring first -- nothing
sent'), dry run included, instead of sending a write the unit ignores.
And the block must be the call-home block: micromate.writes.is_call_home_content
checks its header is its own length (00 7C, N = 124 -- 00 7C DC on every read,
UM12947 and UM20147). plan_call_home_enabled refuses any other 126 bytes, the
change refuses before sending, and call_home_state says why (enabled = None).
At the bench this was checked by eye before the monitoring write.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Measured at the bench 2026-11-03 (BE9441): while monitoring, a MiniMate answers
the 0x2C call-home read with its DEVICE-INFO block ('Instantel', 'MiniMate Plus',
its serial) -- as it answers 1A -- and idle again, minutes later, the call-home
block. SFM writes back only a block it just read, so a monitoring MiniMate's
call-home cannot be changed; the change core already refused (byte 5 read as
'unknown(0x6E)', nothing sent), but said so cryptically.
- minimateplus.protocol.is_call_home_block(raw): the block starts with its own
length 0x7C and is N (or N + 2) bytes unescaped; the device-info block fails it.
- The Series-3 change raises UnitStateUnknown (409) with 'did not return its
call-home settings -- a MiniMate answers the call-home read with its
device-info block while monitoring; stop monitoring first ... nothing sent',
for a dry run too. call_home_state returns enabled = None with that reason.
- The page's callHomeRows shows the reason in place of 'Unknown (byte 5 = ...)'.
A blind write while monitoring stays untested on purpose.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
K441LA17.IN0W's report says Units 'in/s and psi(L)' yet prints
MicTriggerLevel '125.0 dB(L)'. The value states its own unit, so that
is the one compared; the report's Units is the fallback. It was the
only MicTriggerLevel mismatch in DL2 (SFM's 0.005158 psi is 125.0 dB).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Group A of the file-only decode, every field matched against every DL2
report (2026-11-03):
- Identity block before the setup name, what the live read gets from
SUB FE: serial k-202, firmware k-147, DSP version k-138, calibration
date k-132 + 'by' k-128, manufacturer k-94, model k-72, firmware
minor k-193. Event.device_info (DeviceInfo gains calibration_date /
calibrated_by). Serial, Version and Calibration 69,983 / 69,983.
- Battery: uint16 hundredths of a volt at k-190 (Event.battery_volts).
69,729 exact; the other 254 are Blastware rounding half up.
- Trigger source: the two bytes between the embedded event record's
lead and fe 80, a bitmask (0x01 Tran 0x02 Vert 0x04 Long 0x08 MicL
0x200 Manual): 6,493 / 6,493. Event.triggered_by.
- Trigger settings in Series IV's 0x1A layout: sources bitmask k-19,
stop mode k-31 (0 = Fixed), each channel's float32 level at
label+30 (geo in/s, mic psi), mic type code at MicL+4 (2 = Linear
Weighting). Geo level 6,447 / 6,447, mic 4,403 / 4,403; Blastware
prints a level when its channel is a trigger source and the event
was not Manual. The mic level's dB uses 6895 Pa/psi (4,272 / 4,272
against 4,271 at 6894.757).
- .h5 attrs for all of them (additive).
- Checker: SFM's side reports the ten fields; the mic trigger level
follows the report's dB/psi. Carve-out
units-is-a-display-preference: dB or psi is Blastware's PC setting
when it wrote the report -- DL2's reports switch fleet-wide in 61
clean runs by write time, and no byte of the file separates them.
Two tests that pinned BatteryLevel / Calibration as blank now pin the
match.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
CLAUDE.md 'Where things stand' now says plainly how prod uses a paired
Blastware ASCII report (save_imported_bw overlays its peaks, notes,
sample rate and record time onto SFM's decode), that only report-less
events run on the decoder alone, and that the overlay is scaffolding to
retire at step 3 of the plan. The checker status is brought up to dev:
18 of 88 fields, the remaining groups A-F in the proposed order.
README roadmap: the same framing and progress.
Protocol reference:
- D.3.3: the footer is the file's last 26 bytes; footer[18:24] is not
fixed (retracted).
- D.6.1: STRT[18] picks the codec.
- D.6.3: the job number.
- D.6.5: the PVS^2 high word is non-zero above 1.28 in/s (retracted
'reads 0000'); the peak/PVS consistency check; Blastware's Number of
Intervals is the monitoring time / the interval.
- D.6.7: refreshed.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Brian's account: units come back from calibration with the internal
battery unplugged, so a comms-test trigger before the date is set
carries the default. DL2 holds 17 such events on 6 units, all within
1995-09-01 12:02 .. 09-03 11:24.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Every DL2 file ends in a valid footer: 0e 08 and two valid times on
70,041 / 70,041. The decoder took the first '0e 08 + plausible year'
instead, and 14 files also hold an earlier record, ~1.1 KB before the
end, with the stop time followed by zeros:
- a histogram started at its own stop -- a day late on BE9558's daily
histograms (K558L9XP.FK0H: 2025-10-08 19:10:01 for Blastware's
2025-10-07 19:14:08); T003LCYV.JD0H got no time at all;
- a waveform lost its stop, so the trigger was never computed and the
stop was stamped instead (K440I4LG.JK0W: 11:14:12 for 11:14:08).
The last 26 bytes are now taken when they are a footer; otherwise the
old search stands unchanged.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
- s4_change_call_home's read_state now returns an unknown(0xNN) run state as
it is, so the change core refuses it (UnitStateUnknown, nothing sent), as
start/stop already does; it checked only for calling_home.
- run_change records the snapshot on the lost-ack path too: after an ack is
lost, what the unit now holds (and, on Series 4, whether [118:120] moved) is
what tells whether the write landed.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
call_home_image returns the unescaped N + 2 image, but the send path handed it
to write_call_home_config, whose write_image unescapes again -- and unescaping
is not idempotent: a real 0x10 before 02/03/04 loses its 0x10. A call-in time
of 16:02-16:04 (hour 0x10) is exactly that. One such time made the send refuse
before the wire yet audited as 'sent, acknowledgement lost'; two made it send a
block shifted by two bytes from [101] (num_retries and the time fields wrong).
The Blastware-frame test could not see it: it checked the image builder, not
the bytes the protocol framed.
- MiniMateProtocol.write_call_home_image(image) frames a built image once
(0x7E, ack 0x81, 0x7F, ack 0x80), raising ImageRefused before any byte when
it is not a whole block; write_call_home_config (the legacy multi-field path)
is write_image once + that. Its stale 127-byte docstring is gone.
- MiniMateClient.write_call_home_image uses it.
- The Series-3 change passes is_refusal=ImageRefused, so a pre-wire refusal is
audited as nothing sent (sent = 0), not as a lost ack.
Tests drive the real client -> protocol over a stub answering with BE11529's
recorded acks: the unit's own read, a 16:02 time, and 16:03 + 16:04 all go out
exactly as built; a short image is refused with nothing written.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Two test-side breaks, found by the full suite after phase-4 task 6:
- test_micromate_framing's three whole-session tests pinned exact counts of a
gitignored corpus (bridges/captures/9-24-26 - micromate2, a symlink into the
main checkout). The full copy of the BW VM's captures on 2026-11-02 added
sessions: 287 responses / 241 reads / 287 requests, every one passing the same
per-frame checks. The counts are now floors (>= 251 / 218 / 251), so a lost
file still fails and a grown corpus does not.
- test_micromate_opcodes' beyond-the-ceiling permit took 0x7E, enableable since
the call-home commit (eeb55ef); 0x71 (the setup write, phase 5) replaces it.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
device_routing.js (Node-tested): call_home_state / call_home_on / call_home_off
routes (serial required); controlsFor gains callHomeChange (by serial, a
confirmed Series 3 or 4) and callHomeRead now covers a Series 4 by serial;
serialPanes gains callHomeForm / callHomeBox; callHomeRows renders the switch
(On / Off / 'Unknown (byte 5 = 0xNN)'), dial string and series;
changeErrorText takes the whole label ('Start monitoring', 'Enable call-home').
sfm_webapp.js: by serial the Call Home tab hides the legacy multi-field form
(Series 3 by host only) and shows the unit's switch with On / Off. Each asks
confirm(), sends one audited change, reports it with changeSummary /
changeErrorText, then re-reads quietly so the change's own result stays on
screen. A different unit's switch is never left showing. The new buttons are
wired with addEventListener, not onclick= (no inline JS).
Checked in jsdom against the real page: placeholder, read, Off (confirm, the
disable route, 'confirmed -- enabled -> disabled'), a serial edit clearing it,
legacy mode unchanged, no script errors.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- DeviceAccess.call_home_state (a read) and set_call_home (change=True: never
retried, audit required for a live change, dry run without one), routed by
series like start/stop. A config write is read back on its own window: up
to 3 reads, 1 s apart, not the monitoring one.
- Routes, serial required (422 without), all three serial-aware: the state as
a dict; enable / disable through _change_response, so their statuses are
start/stop's and an unconfirmed change carries sent + audit_id. The legacy
GET / POST /device/call_home are unchanged (a Micromate serial is still
refused there, 501).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Ground truth: Blastware changing BE11529's call-home settings on 2026-04-20
(Remote Access -> Setup Unit), paired raw files of what it sent and what the
unit answered: the read, the enable (read 0x00 -> wrote 0x01), the disable
(read 0x01 -> wrote 0x00). From the unit's own read before each write, SFM's
0x7E frame (call_home_image + build_bw_write_frame) is Blastware's byte for
byte -- nothing but byte 5 differs from the read -- and the 0x7F confirm too.
Mutation-checked: inverting the byte SFM writes fails the test.
S3_CALL_HOME_FRAME_VERIFIED is now True; the gate stays tested shut by
monkeypatch, and a test pins that its comment names this proof.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- minimateplus.protocol: call_home_enabled(raw) reads byte 5 of the UNESCAPED
0x2C payload (0x01 on / 0x00 off, measured 2026-04-20; anything else None --
bench finding #1's lesson: an escaped byte before it would move the raw
index); call_home_image(raw, enabled) is write_image's N + 2 image with byte 5
set and everything else as read, refusing a byte 5 never measured.
- MiniMateClient.read_call_home_raw() / write_call_home_image(image) (0x7E then
0x7F, each ack checked, never retried).
- Series3Adapter.call_home_state (wake, then 0x2C; no serial read, like status)
and change_call_home: the shared change core with a connect() identity check,
state 'enabled' / 'disabled' / 'unknown(0xNN)' (unknown sends nothing), the
plan showing both exact frames, a snapshot of both unescaped images.
- S3_CALL_HOME_FRAME_VERIFIED = False: a live change is refused (501) before any
connection until the frame is reproduced from Blastware's (next commit).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- micromate: ENABLEABLE_WRITES gains 0x7E/0x7F (the call-home write and its
confirm; spec row 9, enable/disable only); MicromateClient.get_call_home_content()
returns the 126-byte 0x2C content.
- sfm/device_access: CallHomeState (serial, series, enabled, dial_string, detail
with byte5 and the whole block); s4_call_home_state reads bit 0 of [5] and the
dial string at [6:46]; s4_change_call_home runs the shared change core with
micromate.writes.plan_call_home_enabled -- built from the read run_change made
just before, and that same plan is the one sent. A unit calling home (run
state 0x03) is refused, nothing sent. The read-back judges the edited bit
only; the snapshot records both blocks and every changed offset -- [118:120]
is unit-maintained and always moves.
- The write-ceiling tests take 0x71 (the setup write, phase 5) as their
over-the-ceiling example now that 0x7E is enableable.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
run_change named its change and its target state from a start/stop bool. A
ChangeKind (action, want, became) now says what the change is called and which
state read back means it took; monitoring_change(start) gives exactly today's
names, call_home_change(enabled) the new ones. An optional snapshot hook
records what the unit holds after the change in the result and the audit row
(never the verdict; one that raises is recorded as its error).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
- #6 Config tab by serial: the empty legacy edit form stays hidden (settings
are read-only until phase 5); the settings box shows from the start, 'Not read
yet', and a different unit's settings are never left on screen.
- #10 A start/stop that went out but was not confirmed is no longer called
'failed'. The route now adds 'sent' (acknowledged / unknown) and 'audit_id'
to a ChangeUnconfirmed -- same status, same message -- and the page says
'Start monitoring not confirmed: <what happened, audit record>'.
- #12 Device tab by serial: the empty legacy Compliance/Project tables are
hidden behind a pointer to the Config tab; a Micromate's Firmware card reads
'Thor line (version not read)' (its version string is in 0x01, undecoded).
- #16 A 504 with nothing sent adds: the connection opened, so the modem is
reachable -- another program (THOR or Blastware) may be holding the line, or
the unit is busy. Measured 2026-10-29: THOR polling leaves SFM a 0-byte 504.
New logic lives in sfm/static/device_routing.js (firmwareLabel,
deviceErrorText, changeErrorText, serialPanes) under Node tests; the page glue
in sfm_webapp.js only applies them. Checked in jsdom against the real page
(served at a fake origin, /static from disk): all four, legacy mode unchanged,
no script errors.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
Three causes behind the histograms the checker found decoding wrongly:
1. The file decoder tried the waveform codec first, whatever the record
type. It accepted BE9440 / BE9557 / BE17354's 2018 5 s histograms and
returned ~5x too many 'samples' (K440HDXX.XH0H: SFM 44,504, Blastware
6,976). The codec now follows STRT[18] (0x46 / 0x0A on 70,041 /
70,041 DL2 files); only an unknown type falls back to trial order, and a
chosen codec that fails is a decode problem rather than a silent switch.
2. The sub-minute layout check required each record's 'zero word' to be
0000. It is the high word of the interval's PVS squared, non-zero above
1.28 in/s, so a one-block histogram with a loud interval was not
recognised (T193KR5M.X90H, K440HA3C.JM0H, R352HFZ7.530H decoded to
nothing). The check is now the measured relation peak <= PVS <=
sqrt(3) * peak, which holds on 34,502,955 / 34,502,955 DL2 sub-minute
intervals. The no-terminal-tail trim uses the same test. Two test
builders wrote PVS bytes no real record has (one still in the retracted
little-endian layout); they now write peak^2 in the measured layout, and
the retraction test keeps a legacy builder of its own.
3. Checker carve-out interval-count-is-duration: Blastware's Number of
Intervals is (stop - start) / interval from the footer times, and its
own interval table lists one block fewer when the unit never wrote the
last minute. SFM matches the table (T193KSWP.FF0H: 2,880 vs 2,850).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
Stack-wide rule since 2026-11-02 (terra-view docs/tmi-stack.md, Terra-View's
rule 4): no inline JavaScript, because a test cannot load it. This page's tests
had to regex their code out of the HTML. Batch 2 edits both of the page's
scripts, so they come out first, verbatim:
- sfm/static/device_routing.js: the tested routing rules (157 lines)
- sfm/static/sfm_webapp.js: the rest of the page (2,274 lines)
The page loads them with <script src="static/..."> where the inline scripts
were, so load order and the onclick handlers are unchanged. Verified: putting
the files back inside those tags reproduces the old sfm_webapp.html byte for
byte, and a jsdom page check (served at http://sfm.local/, /static from disk,
no network) gives identical output before and after.
The server serves sfm/static/ at /static/, with Cache-Control: no-cache so a
browser revalidates (a cheap 304) and a new SFM never runs a cached old
script. Opened from disk, the page needs its static/ folder beside it.
Tests: the routing tests load the file instead of regexing the HTML; new
test_webapp_static pins no inline <script> in the page and the /static route
(through the whole app, middleware included).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01SyJ2ip3Krb6dtdv2cXua44
The setup block has Series IV's 0x1A layout here too: report flags at
k - 9 (bit 0x02 = job number shown) and the number, uint16 BE, at k - 8,
kept while switched off. The flag matches Blastware's JobNumber on
2,997 / 2,997 DL2 events that print one and 66,986 / 66,986 that don't.
Every printed value is 1 and matches; hidden ones hold real numbers
(802, 1012, 1013, 622), so the value is only proven for 1.
- ProjectInfo.job_number_enabled / job_number, named as in
micromate.setup_block.
- write_event_hdf5 stores both as .h5 attrs when read (additive).
- The checker reports the number only when shown, and scores a number
only SFM shows as a mismatch. The old test that pinned JobNumber as
sfm_blank now checks only the ZC cases it also covered.
Not carried into the DB, the API or the PDF report yet (needs a column).
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9
The pre-trigger float in an event file's setup block varies from event
to event on the same unit, often on the same day (BE18193: -0.25 and
-0.248 on 2026-08-21; 253-256 samples at 1024 sps, -0.002 to -0.249 s
at 4096 sps), and Blastware reports the same per-event value as
Pre-triggerLength. So it is the event's own pre-trigger, and the note
written earlier today overstated what the files say about #48: the live
1A holds the setting, so a missing anchor in a file says little about
whether the live read loses its anchor. #48 stays open.
Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01ALX2R8pghDaZuti9U7mEE9